在成年人中使用可穿戴的Sensor Dot使用3Hz尖峰波放电的家庭记录缺席
Lauren Swinnen1, Christos Chatzichristos2, Miguel Bhagubai2
1Laboratory for Epilepsy Research, KU Leuven, Leuven, Belgium.
Epilepsia
|November 30, 2023
概括
一个可穿戴的Sensor Dot设备在没有患者在家的情况下准确监测了3Hz的尖峰波放电 (SWD). 与传统日记相比,这项技术提供了更好的扣押记录,患者接受度很高.
科学领域:
- 神经学和生物医学工程
- 监测和可穿戴技术
背景情况:
- 耐药性缺席性的管理依赖于准确的发作数量,经常被不可靠的患者日记所取代.
- 对3赫兹尖峰波放电 (SWD) 的客观家庭监测可以增强对缺席的临床护理.
研究的目的:
- 评估Sensor Dot可穿戴设备的可用性和性能,用于在没有的情况下在家监测SWD.
- 为了比较客观的可穿戴数据与主观的患者日记.
主要方法:
- 13名患有缺席的参与者使用传感器点在家中记录双通道EEG,加速度计和陀螺仪数据.
- 一个自动化算法使用EEG标记了潜在的3HzSWD,并采用多式联接后处理,将运动数据纳入以减少工件.
- 性能指标 (灵敏度,精度,F1得分) 根据检测到的SWD的专家审查进行计算.
主要成果:
- 传感器点在394小时的记录中捕获了11名参与者的234个SWD.
- 单模算法实现了0.84的平均灵敏度,0.93的精度和0.89的F1得分.
- 多模式分析显示了可比性能,审查时间缩短;患者发现设备舒适和可接受.
结论:
- 传感器点为缺席提供了比患者自我报告更准确的家庭文档.
- 该设备的用户友好性,舒适性和短的审查时间有助于其临床实用性和患者接受.
更多相关视频
08:43Long-term Continuous EEG Monitoring in Small Rodent Models of Human Disease Using the Epoch Wireless Transmitter System
Published on: July 21, 2015
25.6K
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
8.7K
相关概念视频
Pulse rhythm
1.7K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.7K
Equipments Used To Measure Blood Pressure
3.4K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.4K
Holter Monitor: 24-Hour Monitoring
5.7K
Holter monitoring is a continuous electrocardiography (ECG) recording that tracks the heart's electrical activity over an extended period, generally 24 to 48 hours. This noninvasive diagnostic tool detects irregular heart rhythms that may not be captured during a standard ECG performed in a clinical setting.DeviceThe Holter monitor is a portable, small device connected to several electrodes on the patient's chest. These electrodes detect the heart's electrical signals and transmit them to the...
5.7K
Electronic Distance Measuring Instruments
771
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over...
771