圆柱形离子陷的实证校准用于大规模选择的气相光谱学
Thomas Toft Lindkvist1, Jeppe Langeland1, Christina Kjær1
1Department of Physics and Astronomy, Aarhus University, DK-8000 Aarhus C, Denmark.
Journal of the American Society for Mass Spectrometry
|December 11, 2023
概括
我们开发了一种使用光检测的圆柱形离子陷的新校准方法. 这种技术准确地绘制了离子稳定性,改善了它们作为分子离子质谱仪的使用.
科学领域:
- 分析化学 分析化学
- 物理化学 物理化学
- 原子和分子物理 原子和分子物理
背景情况:
- 四极离子陷使用马修的方程来分析离子运动.
- 圆柱形离子陷提供了诸如用于光谱学的增强光子采集等优势.
- 了解圆柱状陷中的离子动力学是质谱应用的关键.
研究的目的:
- 介绍一种使用光检测的圆柱形离子陷的实证校准方法.
- 在实验条件下分析圆柱形离子陷内的离子运动和稳定性.
- 为了确定圆柱形离子陷的质谱能力.
主要方法:
- 在脉冲实验中通过光检测对圆柱形离子陷的实证校准.
- 应用有效的马修描述与有效的几何参数用于陷中心的离子运动.
- 六个分子离子 (四个阴离子,两个阴离子) 的映射稳定性图,跨越m/z 212-647.7.
主要成果:
- 实现了几乎没有背景的校准方法.
- 圆柱形离子陷的稳定性图与马修图对齐,由碰撞转移.
- 稳定图的参数是使用具有有效几何和q依赖参数的马修函数参数化.
- 质量分辨率估计为数百个Da范围内的离子的+7/-3Da.
结论:
- 经验校准方法提供了一种可靠的方式,以了解圆柱形离子陷中的离子行为.
- 圆柱形离子陷,尽管来自缓冲气体的复杂性,显示可预测的离子稳定性与马修预测相比.
- 这项工作验证了使用圆柱形离子陷作为一系列分子离子的有效质谱仪.
相关概念视频
Mass Analyzers: Common Types
615
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
615
Gas Chromatography: Types of Detectors-II
381
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
381
Gas Chromatography: Types of Detectors-I
429
There are different types of detectors used in gas chromatography, each with its own specific properties that make it suitable for detecting certain types of analytes. The most commonly used detectors in GC are thermal conductivity detector (TCD), flame ionization detector (FID), and electron capture detector (ECD).
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
429
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
749
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
749
Mass Analyzers: Overview
683
The mass analyzer is a crucial component of the mass spectrometer. In the ionization chamber, the vaporized sample is bombarded with a high-energy electron beam to generate a radical cation and further fragment into neutral molecules, radicals, and cations. A series of negatively charged accelerator plates accelerate the cations into the mass analyzer. The mass analyzer separates ions according to their mass-to-charge (m/z) ratios and then directs them to the detector. The common types of mass...
683
Atomic Emission Spectroscopy: Lab
167
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
167