概括
深度大脑刺激 (DBS) 导致的交叉声会显著影响MRI安全性. 由于这种相互作用,DBS导尖的功率沉积可能会变化多达六倍,需要进一步的安全评估.
科学领域:
- 医学物理 医学物理
- 神经外科 神经外科
- 生物医学工程 生物医学工程
背景情况:
- 深度大脑刺激 (DBS) 是帕金森病和其他神经系统疾病的关键治疗方法.
- 目前对DBS植入物进行的MRI安全评估是单独评估线索,忽视线索间的相互作用 (交叉声).
研究的目的:
- 为了研究双边DBS线索之间的交叉声波对MRI期间射频 (RF) 功率沉积的影响.
- 评估配置如何影响射频功率沉积和潜在的热风险.
主要方法:
- 从CT图像获得患者特定的轨迹,用于数值模拟.
- 在单边和双边DBS植入场景中,将尖的射频功率沉积量进行了比较.
- 为了验证模拟结果,使用简化的DBS引线系统进行了幻影实验.
主要成果:
- 数字模拟显示,在DBS引线尖端的射频功率沉积可以增加多达六倍,双边引线与单边引线相比.
- 实验测量证实了DBS线路之间存在交叉通话的存在.
- 多个导线的存在可以改变能量沉积,可能增加或减少局部组织加热.
结论:
- 对于DBS植入物的MRI安全性评估必须考虑交叉声.
- 双边DBS引线配置可以显著影响射频功率沉积,影响MRI扫描期间的患者安全.
- 需要进一步的研究来完善双边DBS植入物患者的MRI协议.
相关概念视频
Magnetic Resonance Imaging
5.2K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
5.2K
Double Resonance Techniques: Overview
214
Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
Spin decoupling is usually achieved by...
214
Atomic Nuclei: Nuclear Relaxation Processes
657
In the absence of an external magnetic field, nuclear spin states are degenerate and randomly oriented. When a magnetic field is applied, the spins begin to precess and orient themselves along (lower energy) or against (higher energy) the direction of the field. At equilibrium, a slight excess population of spins exists in the lower energy state. Because the direction of the magnetic field is fixed as the z-axis, the precessing magnetic moments are randomly oriented around the z-axis.
657
Atomic Nuclei: Magnetic Resonance
663
The number of nuclear spins aligned in the lower energy state is slightly greater than those in the higher energy state. In the presence of an external magnetic field, as the spins precess at the Larmor frequency, the excess population results in a net magnetization oriented along the z axis. When a pulse or a short burst of radio waves at the Larmor frequency is applied along the x axis, the coupling of frequencies causes resonance and flips the nuclear spins of the excess population from the...
663
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
870
An applied magnetic field causes the electrons present in the molecule to circulate, setting up a local diamagnetic current within the molecule. The local diamagnetic current arising from circulating sigma-bonding electrons induces a magnetic field, Blocal that opposes the applied magnetic field, B0. The effective magnetic field experienced by these nuclei is given by the difference between the applied and local magnetic fields in a phenomenon called local diamagnetic shielding. Essentially,...
870