对于9.4 T全身MRI超导磁铁的被动闪
Wenchen Wang1, Hongyi Qu2, Yaohui Wang2
1School of Information Technology and Electrical Engineering, University of Queensland, Brisbane, QLD 4072, Australia.
The Review of scientific instruments
|December 14, 2023
概括
研究人员开发了一种800毫米热孔超导磁铁,用于9.4 T全身MRI. 被动闪实现了高磁场均性,这对于先进的磁共振成像应用至关重要.
科学领域:
- 医学物理 医学物理
- 超导磁铁技术 超导磁铁技术 超导磁铁技术
- 磁共振成像是一种磁共振成像技术.
背景情况:
- 超高场 (UHF) 超导磁铁的开发对于推进全身磁共振成像 (MRI) 系统至关重要.
- 实现高磁场均性对于在MRI中获得高分辨率图像至关重要.
研究的目的:
- 详细介绍了对9.4 T,800 mm热孔MRI超导磁铁的被动闪技术的实施.
- 分析超高场MRI磁铁中与被动闪相关的电磁力和系统错误.
主要方法:
- 设计和实施9.4 TMRI超导磁铁的被动闪.
- 使用代调整来优化磁场在球体体积 (DSV) 的400毫米直径上的同质性.
- 对超高频MRI系统中被动闪特有的电磁力和系统错误的分析.
主要成果:
- 在DSV上实现了3.05 ppm的峰值到峰值磁场均性.
- 在DSV上实现了0.94ppm的根平均平方 (RMS) 场均质.
- 在超高频MRI磁铁中被动闪的量化电磁力和系统错误.
结论:
- 被动闪的成功实施显著改善了9.4 T MRI磁铁的磁场均性.
- 该研究为超高场MRI磁铁开发的被动闪提供了有价值的工程见解.
- 开发的磁铁和闪技术为增强的MRI能力铺平了道路.
相关概念视频
Magnetic Resonance Imaging
5.2K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
5.2K
Double Resonance Techniques: Overview
214
Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
Spin decoupling is usually achieved by...
214
Magnetism
6.4K
Magnets are commonly found in everyday objects, such as toys, hangers, elevators, doorbells, and computer devices. Experimentation on these magnets shows that all magnets have two poles: one is labeled north (N) and the other south (S). Magnetic poles repel if they are alike and attract if unlike. Moreover, both poles of a magnet attract unmagnetized pieces of iron.
An individual magnetic pole cannot be isolated. No matter how small, every piece of a magnet contains a north pole and a south...
An individual magnetic pole cannot be isolated. No matter how small, every piece of a magnet contains a north pole and a south...
6.4K