作为反射脉冲氧计的Empatica EmbracePlus腕带的未来临床验证
Giulia Gerboni1, Giulia Comunale1, Weixuan Chen1
1Empatica, Inc., Cambridge, MA, United States.
在受控缺氧期间,EmbracePlus腕带准确地测量了不同个体的外周氧和 (SpO2). 这种可穿戴设备显示了持续氧气监测的临床有效性,有助于早期检测低氧症.
科学领域:
- 生物医学工程 生物医学工程
- 心肺生理学 心肺生理学
- 医疗器械技术 医疗器械技术
背景情况:
- 呼吸系统疾病可以降低动脉氧和度 (SaO2),需要持续监测以检测低氧症.
- 可穿戴的脉冲氧度仪器提供非侵入性的外周氧和 (SpO) 监测.
- 临床验证至关重要,特别是在不同种族群体的表现方面.
研究的目的:
- 为了对EmbracePlus腕带反射脉冲氧计功能的准确性进行前性验证.
- 根据ISO 80601-2-61:2017和FDA指南,在受控的低氧条件下评估设备的性能.
- 为了评估不同皮肤颜色和性别人口统计数据的准确性.
主要方法:
- 健康的成年参与者接受了受控的脱协议 (SaO2100%到70%).
- 将EmbracePlus腕带SpO2的测量与来自同氧度血液分析的参考SaO2进行了比较.
- 数据包括30%的菲茨帕特里克皮肤类型V-VI的参与者.
主要成果:
- 该研究实现了2.4%的精度平方根平均值 (rms) 误差,低于FDA的3.5%限制.
- 在SpO2和SaO2之间观察到强烈的正相关性 (r=0.96) 和良好的一致性 (bias 0.05%,SD 1.66%).
- 准确度在菲茨帕特里克皮肤类型 (V-VI的2.2%) 和性别之间仍然可以接受.
结论:
- 在特定条件下,EmbracePlus腕带在不同个体的SPO评估中表现出临床上可接受的准确性.
- 这些发现支持其在各种人群中用于持续氧气监测的潜在用途.
- 需要在不健康的受试者和门诊环境中进一步验证.
更多相关视频
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
相关概念视频
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation