Jove
Visualize
联系我们
JoVE
x logofacebook logolinkedin logoyoutube logo
关于 JoVE
概览领导团队博客JoVE 帮助中心
作者
出版流程编辑委员会范围与政策同行评审常见问题投稿
图书馆员
用户评价订阅访问资源图书馆顾问委员会常见问题
研究
JoVE JournalMethods CollectionsJoVE Encyclopedia of Experiments存档
教育
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab Manual教师资源中心教师网站
使用条款与条件
隐私政策
政策

相关概念视频

Imaging Studies II: Positron Emission Tomography and Scintigraphy01:25

Imaging Studies II: Positron Emission Tomography and Scintigraphy

122
Positron Emission Tomography (PET) is a medical imaging technique that provides crucial insights into the body's physiological functions at a molecular level. It is an indispensable resource for diagnosing, staging, and monitoring various illnesses, notably cancer, neurological disorders, and cardiovascular conditions.
Fundamental Principles of PET
122
Radioactive Decay and Radiometric Dating02:48

Radioactive Decay and Radiometric Dating

34.1K
Radioactivity is a spontaneous disintegration of an unstable nuclide and is a random process, as all the nuclei in the sample do not decay simultaneously. The number of disintegrations per unit time is called the activity (A), which is directly proportional to the number of nuclei in the sample. The decay constant (λ) is an average probability of decay per nucleus in unit time.
34.1K

您也可能阅读

相关文章

通过共同作者、期刊和引用图与本文相关的文章。

排序
Same author

Efficacy of peptide receptor radionuclide therapy in patients with oligodiscordant gastroenteropancreatic neuroendocrine tumours.

European journal of nuclear medicine and molecular imaging·2026
Same author

PET/CT contouring techniques in neuroendocrine neoplasms: a state-of-the-art scoping review.

EJNMMI reports·2026
Same author

Somatostatin receptor PET response assessment framework for patients with neuroendocrine tumours (V1.0): a modified Delphi consensus from the European Neuroendocrine Tumor Society (endorsed by EANM and NANETS).

The Lancet. Oncology·2026
Same author

[<sup>177</sup>Lu]Lu-DOTATATE PRRT and CAPTEM chemotherapy for pancreas and small bowel neuroendocrine tumours: The AGITG CONTROL NETS Multi-centre randomized trial.

European journal of cancer (Oxford, England : 1990)·2026
Same author

Biological and comorbidity factors outweigh lymph node metrics in predicting outcomes after small-bowel neuroendocrine tumour resection.

Surgical oncology·2026
Same author

Therapy-related myeloid neoplasms following peptide receptor radionuclide therapy for neuroendocrine neoplasms: case series reporting characteristics and outcomes from single-centre experience.

Endocrine oncology (Bristol, England)·2026

相关实验视频

Updated: Jul 6, 2025

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
08:53

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level

Published on: June 6, 2018

8.0K

用长寿命放射性核素评估定量扫描器准确性的实用方法:ARTnet插件

Dale L Bailey1,2, Kathy P Willowson1,3, Carl Muñoz-Ferrada4

  • 1Department of Nuclear Medicine, Royal North Shore Hospital, Sydney, Australia.

Asia Oceania journal of nuclear medicine & biology
|January 2, 2024
PubMed
概括

一个新的测试对象,ARTnet Insert,通过使用更少的放射性和更小的体积,简化了PET和SPECT系统的校准. 这种新的幻影插件确保了准确的图像重建,没有部分体积效应.

关键词:
临床试验 临床试验在这里,PET是PET.斯佩克特 (Spectre) 是一个运动场.网站验证 网站验证

更多相关视频

A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
14:19

A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space

Published on: February 1, 2016

8.6K
Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases
11:03

Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases

Published on: May 29, 2017

10.0K

相关实验视频

Last Updated: Jul 6, 2025

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
08:53

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level

Published on: June 6, 2018

8.0K
A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
14:19

A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space

Published on: February 1, 2016

8.6K
Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases
11:03

Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases

Published on: May 29, 2017

10.0K

科学领域:

  • 医学成像物理 医学成像物理
  • 核医学 技术 技术 核医学

背景情况:

  • 目前用于校准正子发射断层扫描 (PET) 和单光子发射计算机断层扫描 (SPECT) 系统的方法通常需要在测试幻体中大量的长寿命放射性核素.
  • 这带来了与放射性处理,储存以及潜在的部分体积影响影响准确性相关的挑战.

研究的目的:

  • 开发一种用于PET和SPECT系统校准的新型测试对象,以最大限度地减少放射性体积并简化存储.
  • 为了确保测试对象在没有部分体积效应的情况下保持精度,即使使用较低空间分辨率的成像.

主要方法:

  • 一个空洞的烯酸圆柱形插件,ARTnet插件,是为了与NEMA/IEC PET身体图像质量 (IQ) 幻影相容而设计的.
  • 插件的尺寸 (90毫米内部直径,70毫米内部高度) 是为了减轻部分体积效应而选择的.
  • 在校准扫描仪上用各种放射性核素 (18F,68Ga,124I用于PET/CT;99mTc,131I,177Lu用于SPECT/CT) 进行测试.

主要成果:

  • 在ARTnet插件允许准确的图像重建在5%的真实值.
  • 与NEMA NU 2-2018幻影测量相比,没有观察到部分体积效应的混不确定性.
  • 插入器即使在高能量的SPECT成像 (例如131I) 中也有效,其中空间分辨率通常较差.

结论:

  • ARTnet插件是PET和SPECT校准的简单,廉价和多功能工具.
  • 与传统方法相比,它减少了放射性要求和储存挑战.
  • 它的设计有效地克服了部分体积效应,确保了在不同成像方式和放射性核素中可靠的校准.