这是首次对人类进行评估基于智能戒指的无袖血压测量的准确性研究
Jihoon Kim1, Sung-A Chang1, Seung Woo Park2
1Division of Cardiology, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
Journal of Korean medical science
|January 16, 2024
概括
一种新的环型可穿戴设备可以准确地测量血压 (BP),而无需手套. 这项首次在人体上进行的试验显示了最小的偏差和与传统血压测量方法的强烈相关性.
科学领域:
- 生物医学工程 生物医学工程
- 心血管技术的心血管技术
- 可穿戴的健康设备
背景情况:
- 开发了一种新的环型无袖血压 (BP) 监测设备.
- 这项研究代表了一项前性,单臂,首次在人体中进行的关键试验,以评估这种新型血压测量设备的准确性.
研究的目的:
- 为了评估新型环型无袖血压 (BP) 测量装置的准确性.
- 将环装置的血压测量与参考听力方法进行比较.
主要方法:
- 戒指设备利用来自手指的光电显微镜 (PPG) 信号,通过无线传输到智能手机.
- 同时进行血压测量,使用一只手臂上的戒指装置,另一只手臂上的传统袖口.
- 来自89名受试者的数据 (526个缩血压和513个缩血压组) 被分析为平均差异和相关性.
主要成果:
- 环状装置显示了0.16 ± 5.90 mmHg的平均差异对于静缩血压和 -0.07 ± 4.68 mmHg对于心动血压.
- 观察到环状装置和对缩血压 (r=0.94) 和腹缩血压 (r=0.95) 的参考测量之间存在很强的相关性.
- 结果在样本智和对象智的比较中一致,在可接受的临床范围内达成95%的一致性.
结论:
- 环型血压测量装置与标准听力法相比,具有良好的相关性和最小的偏差.
- 这种可穿戴技术显示出准确,无袖式血压监测的前景.
相关概念视频
Assessment of blood pressure in brachial artery(two-step method)
711
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
711
Pre-Procedural Guidelines for Assessing Blood Pressure
565
Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
565
Special considerations while measuring blood pressure
724
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
724
Sites for measruring blood pressure
1.7K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
1.7K
Equipments Used To Measure Blood Pressure
973
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
973
Assessment of blood pressure in brachial artery(one-step method)
593
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
593