开发一个测量系统的口内SPO2
Yuki Kashima1, Minako Onimaru1, Ryosuke Isogai2
1Department of Orthodontics, School of Dentistry, Showa University, 2-1-1 Kitasenzoku, Ota-ku, Tokyo 145-8515, Japan.
Sensors (Basel, Switzerland)
|January 23, 2024
概括
这项研究引入了一种新的口内测试方法,用于使用可穿戴设备测量血液中氧和度 (SpO2). 与基于手指的方法相比,口内SPO2监测可以更快地检测变化.
科学领域:
- 生物医学工程 生物医学工程
- 医疗器械 医疗器械
- 身体生理学 身体生理学
背景情况:
- 血中氧和度 (SpO2) 是一个关键的患者生命体征.
- 传统的SpO2测量方法面临着时间延迟和环境光干扰等挑战.
- 没有先前的研究存在于使用可穿戴技术的口内SPO2测量.
研究的目的:
- 开发和验证一种新的口内SPO2测量技术.
- 为了确定口内SPO2传感的最佳条件.
- 为了比较口内与指尖 SpO2 监测的响应性.
主要方法:
- 在上六个口腔内位置评估输液指数 (PI),以找到最佳位置.
- 确定PI验证的最佳测量压力 (0.30.8 N).
- 在12名健康成年人中,使用定制的口罩设备在呼吸中同时进行口内和指尖的SpO2测量.
主要成果:
- 在大牙附近的口腔牙产生了最高的PI值.
- 在0.3至0.8N的压力范围内,可以达到最佳PI.
- 由于靠近心脏,口内SPO2变化比指尖测量快大约7秒.
结论:
- 口腔是使用新型设备获取生物信息的可行位置.
- 与传统方法相比,口内SPO2监测显示出更快的反应时间.
- 这种新型的可穿戴设备为改善SpO2监测提供了潜力.
相关概念视频
Guidelines For Measuring Vital Signs
1.6K
Following these guidelines can help nurses accurately measure vital signs, assess changes in patient conditions, and provide timely treatment when necessary. Adhering closely to the guidelines ensures the accuracy and reliability of the results.
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
1.6K
Pulse Oximetry
329
Pulse oximetry, or SpO2, is a non-invasive method for continuously monitoring arterial oxygen saturation (SaO2). This procedure involves attaching a probe or sensor to the patient's fingertip, forehead, earlobe, or nose bridge. The sensor works by detecting changes in oxygen saturation levels through light signals generated by the oximeter and reflected by the pulsing blood under the probe.
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
329
Special considerations while measuring oxygen saturation
592
Assessing respiratory rate concurrently with pulse measurement is fundamental to patient care, providing valuable insights into the patient's respiratory function. The normal breathing rate for an adult usually falls within a normal range of 12 to 20 breaths per minute. Abnormal respiratory rates can signal underlying health conditions or the need for immediate intervention.
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
592
Equipments Used To Measure Blood Pressure
973
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
973
Temperature Measurement Sites
1.7K
A thermometer measures body temperature. The common sites for measuring body temperature are the oral cavity, axillary region, temporal artery, and skin surface, such as the forehead, abdomen, and axilla. True core body temperature is assessed in the rectum, tympanic membrane, pulmonary artery, esophagus, and urinary bladder.
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
1.7K
Assessing Body Temperature - Oral
745
Here are the steps to accurately measure oral temperature using an electronic thermometer:
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
745