关于开源和参考工业设备之间的一致性和相似性的正式评估,其中包括对低成本pH记录器的应用
Evmorfia P Bataka1, Persefoni Maletsika2, Christos T Nakas1,3
1Laboratory of Biometry, Department of Agriculture, Crop Production and Rural Environment, University of Thessaly, 384 46 Volos, Greece.
Sensors (Basel, Switzerland)
|January 23, 2024
概括
这项研究开发了一种低成本的开源pH记录器,用于农业应用. 在重新校准后,该设备与专业计量仪表显示出极好的一致性,提高了测量精度.
科学领域:
- 农业科学 农业科学
- 仪器化 仪器化 仪器化
- 开源技术是开源技术.
背景情况:
- 开源设备在各种研究领域越来越重要.
- 准确的现场pH监测对于农产品制造至关重要.
研究的目的:
- 设计和开发一个便携式的,低成本的,开源的pH记录器,用于农业实地测量.
- 评估开发的设备与商用pH计之间的一致性.
主要方法:
- 使用标准pH缓冲器 (4.01,7.01) 进行手动校准.
- 包括一个防水温度传感器用于pH补偿.
- 使用混合效应模型和协议指数进行正式方法比较.
主要成果:
- 观察到初始平的协议与0.22pH单位的固定偏差.
- 再校准显著改善了协议的优异水平.
- 拟议的比较方法揭示了仅依赖于相关系数的局限性.
结论:
- 在重新校准后,开发的开源pH记录器为农业应用提供了准确的测量.
- 正式的方法比较对于验证新工具和避免误导性结果至关重要.
- 该研究主张强有力的协议评估超出了简单的相关性.
更多相关视频
08:25Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
2.6K
08:06The Use of a β-lactamase-based Conductimetric Biosensor Assay to Detect Biomolecular Interactions
Published on: February 1, 2018
9.0K
相关概念视频
pH Scale
68.9K
Hydronium and hydroxide ions are present both in pure water and in all aqueous solutions, and their concentrations are inversely proportional as determined by the ion product of water (Kw). The concentrations of these ions in a solution are often critical determinants of the solution’s properties and the chemical behaviors of its other solutes. Two different solutions can differ in their hydronium or hydroxide ion concentrations by a million, billion, or even trillion times. A common means of...
68.9K
Potentiometry: Types of Electrodes
657
Reference electrodes serve as a stable reference point for potentiometric measurements, while indicator and working electrodes react to variations in the composition of a solution.
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
657
Indicators
48.7K
Certain organic substances change color in dilute solution when the hydronium ion concentration reaches a particular value. For example, phenolphthalein is a colorless substance in any aqueous solution with a hydronium ion concentration greater than 5.0 × 10−9 M (pH < 8.3). In more basic solutions where the hydronium ion concentration is less than 5.0 × 10−9 M (pH > 8.3), it is red or pink. Substances such as phenolphthalein, which can be used to determine the pH of a solution, are...
48.7K
Amperometry: Overview
550
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
550