概括
我们开发了一种低成本的电气方法来减少共振扫描显微镜中的动. 这种技术显著提高了图像质量,并且可以在没有软件更改的情况下实现高速成像.
科学领域:
- 光学和成像科学科学 光学和成像科学
- 生物医学工程 生物医学工程
- 显微镜技术 显微镜技术
背景情况:
- 响应扫描对于高速和体内激光扫描显微镜至关重要.
- 在共振扫描中的扫描器动会导致图像工件,阻碍高速成像采用.
研究的目的:
- 推出一种实时,廉价,纯电气的方法来抑制扫描仪动.
- 为了证明这种方法可以在没有软件修改的情况下应用于现有的显微镜系统.
主要方法:
- 阶段锁定成像到共振扫描仪的时间.
- 实施一种纯电动动抑制技术.
- 改装一个商用双光子显微镜.
主要成果:
- 实现了超出离散极限以下一个数量级的动抑制.
- 显示了像素动的86%降低和点差功能的15%改善.
- 启用了共振扫描仪,以匹配较慢的仪扫描仪的准确性.
结论:
- 开发的方法显著提高了定量和定性生物成像.
- 该技术为改进现有的激光扫描显微镜系统提供了多功能解决方案.
- 能够实现高速,高精度的成像,对于先进的显微镜应用至关重要.
相关概念视频
Upsampling
237
Managing signal sampling rates is essential in digital signal processing to maintain signal integrity. A decimated signal, characterized by a reduced frequency range due to its lower sampling rate, can be upsampled by inserting zeros between each sample. This upsampling process expands the original spectrum and introduces repeated spectral replicas at intervals dictated by the new Nyquist frequency. To refine this zero-inserted sequence, it is passed through a lowpass filter with a cutoff...
237
Aliasing
136
Accurate signal sampling and reconstruction are crucial in various signal-processing applications. A time-domain signal's spectrum can be revealed using its Fourier transform. When this signal is sampled at a specific frequency, it results in multiple scaled replicas of the original spectrum in the frequency domain. The spacing of these replicas is determined by the sampling frequency.
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original...
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original...
136
Sampling Continuous Time Signal
251
In signal processing, a continuous-time signal can be sampled using an impulse-train sampling technique, followed by the zero-order hold method. Impulse-train sampling involves the use of a periodic impulse train, which consists of a series of delta functions spaced at regular intervals determined by the sampling period. When a continuous-time signal is multiplied by this impulse train, it generates impulses with amplitudes corresponding to the signal's values at the sampling points.
In the...
In the...
251
Reconstruction of Signal using Interpolation
197
Signal processing techniques are essential for accurately converting continuous signals to digital formats and vice versa. When a continuous signal is sampled with a period T, the resulting sampled signal exhibits replicas of the original spectrum in the frequency domain, spaced at intervals equal to the sampling frequency. To handle this sampled signal, a zero-order hold method can be applied, which creates a piecewise constant signal by retaining each sample's value until the next...
197
Double Resonance Techniques: Overview
210
Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
Spin decoupling is usually achieved by...
210
Bandpass Sampling
181
In signal processing, bandpass sampling is an effective technique for sampling signals that have most of their energy concentrated within a narrow frequency band. This type of signal is known as a bandpass signal. The key principle of bandpass sampling involves sampling the signal at a rate that is greater than twice the signal's bandwidth to prevent aliasing.
A bandpass signal has a spectrum with a lower frequency limit, denoted as ω1, and an upper frequency limit, denoted as ω2....
A bandpass signal has a spectrum with a lower frequency limit, denoted as ω1, and an upper frequency limit, denoted as ω2....
181