基于智能手机的与非侵入性的自动振荡计手臂袖子血压测量:一个前性的方法比较志愿者研究
Lila Delmotte1, Olivier Desebbe2, Brenton Alexander3
1Department of Anesthesiology, Erasme University Hospital, Université Libre de Bruxelles, 808 Route de Lennik, 1070 Brussels, Belgium.
Journal of personalized medicine
|January 26, 2024
概括
与手套相比,OptiBPTM智能手机应用程序在估计血压 (BP) 中显示了可接受的准确性. 虽然符合ISO标准的平均和透气血压,但缩血压需要改进,以便在未来使用.
科学领域:
- 生物医学工程 生物医学工程
- 医疗器械 医疗器械
- 数字健康数字健康
背景情况:
- 移动健康诊断对于慢性疾病管理是有效的.
- OptiBPTM技术使用智能手机的光学信号来估计血压 (BP).
研究的目的:
- 在四周的时间内评估OptiBPTM设备的准确性和精度.
- 为了比较OptiBPTM BP测量与标准的振荡计手臂袖口.
主要方法:
- 没有心律失常的成年参与者在四周内同时使用OptiBPTM和手臂袖口进行血压测量.
- 使用Bland-Altman和错误网分析来比较BP值.
- 性能是根据国际标准化组织 (ISO) 的定义进行评估的.
主要成果:
- 在65名参与者中,有53人提供了适当的OptiBPTM BP值.
- 平均偏差 ± SD: -1.4 ± 10.1 mmHg (心),0.2 ± 6.5 mmHg (腹), -0.5 ± 6.9 mmHg (平均值). 在此过程中,每小时的平均偏差为1.4 ± 10.1 mmHg (心),0.2 ± 6.5 mmHg (腹),0.5 ± 6.9 mmHg (平均值).
- 100%的测量都属于A (无风险) 和B (低风险) 错误网格区域.
结论:
- 在四周的时间里,OptiBPTM与参考方法表现出可接受的一致性.
- 该设备符合ISO标准的平均动脉压 (MAP) 和透气动脉压 (DAP),但不符合系统动脉压 (SAP).
- 在技术改进之前,OptiBPTM显示出作为未来BP监测工具的潜力.
相关概念视频
Sites for measruring blood pressure
1.7K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
1.7K
Assessment of blood pressure in brachial artery(two-step method)
709
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
709
Assessment of blood pressure in brachial artery(one-step method)
591
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
591
Special considerations while measuring blood pressure
724
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
724
Equipments Used To Measure Blood Pressure
972
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
972
Measurement of Blood Pressure
931
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
931