在植入式心脏转换器-除器中,极性依赖,减少或无输出冲击
Charles D Swerdlow1, Michael Ringle2
1Smidt Heart Center, Cedars Sinai Medical Center, Los Angeles, California, USA.
JACC. Clinical electrophysiology
|January 26, 2024
概括
植入式心脏转换器-除器的冲击可能会由于通过短路的输入而失败. 一个拟议的机制解释了取决于极性的故障,即阴极冲击会产生等离子体,导致设备故障.
科学领域:
- 生物医学工程 生物医学工程
- 材料科学 材料科学 材料科学
背景情况:
- 植入式心脏转换器-除器 (ICD) 可以经历极性依赖的故障.
- 这些故障与高压输入通道的结构问题有关,特别是在阴极冲击期间.
研究的目的:
- 提出一种解释ICD中极度依赖冲击失效的机制.
- 调查这个机制是否适用于其他冲击引起的短路.
主要方法:
- 分析ICD输入结构及其在冲击传递中的作用.
- 电场梯度的理论建模,流体蒸发和冲击期间的电子场辐射.
- 在阴极和阳极冲击期间的条件的比较.
主要成果:
- 输入通道中的液体层与电场梯度相结合,可以在冲击时引起水蒸发.
- 对于正极冲击,来自内针的电子发射会导致电离雪崩和等离子形成,导致短路.
- 阳极冲击导致液气接口的电场较弱,稳定气泡,防止故障.
结论:
- 拟议的机制解释了取决于极性的ICD冲击故障.
- 这种机制也可能适用于电子设备中的其他冲击诱导的短路.
更多相关视频
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
3.3K
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10.5K
相关概念视频
Cardiac Action Potential
1.4K
Cardiac action potentials are essential for proper heart function, enabling the rhythmic contractions needed for adequate blood circulation. Nodal cells and Purkinje fibers, specialized for electrical conduction, generate these action potentials.
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
1.4K
Electrophysiology of Normal Cardiac Rhythm
4.4K
The normal cardiac rhythm is a synchronized electrical activity that facilitates the regular and coordinated contraction of the heart muscle. This process is essential for efficient blood circulation throughout the body. The fundamental elements involved in establishing and maintaining this rhythm include the unique electrical properties of cardiac muscle cells, the sinoatrial (SA) node's pacemaker function, the specialized conducting system, and the ionic mechanisms underlying each phase...
4.4K
Disturbances in Heart Rhythm
957
Arrhythmia or dysrhythmia refers to an abnormal heart rhythm caused by a defect in the heart's conduction system. It can cause the heart to beat irregularly, too quickly, or too slowly, leading to symptoms like chest pain, shortness of breath, and fainting. Factors such as stress, caffeine, alcohol, nicotine, cocaine, certain drugs, congenital defects, diseases, and electrolyte abnormalities can trigger arrhythmias.
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
957
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers
746
Adrenergic stimulation generally impacts cardiac rate and rhythm. Specifically, stimulation of the β-adrenoceptors triggers an increase in intracellular calcium ion influx and pacemaker currents, which may cause arrhythmias. Catecholamines like adrenaline also demonstrate β2-adrenoceptor-mediated hypokalemia, impacting cardiac action potential and disrupting the normal cardiac rhythm. Class II antiarrhythmic drugs are β-adrenoceptor antagonists or β-blockers, which...
746
Clipper Circuit
438
A clipper circuit is a fundamental wave-shaping device that harnesses the unique properties of diodes to alter and control waveform characteristics. This technology is widely used in electronic devices, especially in television and radar communication systems, where it enhances waveform modulation in both transmitters and receivers.
The operation of a clipper circuit can be exemplified by analyzing a dual-clipper configuration setup that integrates two ideal diodes, each paired with a biasing...
The operation of a clipper circuit can be exemplified by analyzing a dual-clipper configuration setup that integrates two ideal diodes, each paired with a biasing...
438
Node Analysis for AC Circuits
320
Consider an angioplasty system featuring a catheter equipped with a turbine, a critical tool for removing plaque deposits from coronary arteries. This intricate medical device operates using a circuit model reminiscent of a dual-node RLC circuit powered by a current-controlled voltage source.
To unravel the complexities of this system, nodal analysis is employed, a powerful technique founded on Kirchhoff's current law (KCL), which remains valid for phasors. AC circuits can effectively be...
To unravel the complexities of this system, nodal analysis is employed, a powerful technique founded on Kirchhoff's current law (KCL), which remains valid for phasors. AC circuits can effectively be...
320