使用基于外围动脉导管的波形来估计心脏输出的Edwards Acumen IQ系统与狗的热稀释相比不准确
American journal of veterinary research
|February 9, 2024
概括
与狗的热稀释相比,Edwards Acumen IQ传感器 (HS-IQ) 系统用于估计心脏输出 (CO) 显示了广泛的协议限制. 目前的HS-IQ CO估计不适合临床应用.
科学领域:
- 心血管生理学心血管生理学
- 医疗器械评估 医疗器械评估
- 麻醉学 麻醉学
背景情况:
- 精确的心脏输出 (CO) 监测对于管理重症患者至关重要.
- 对CO估计的非侵入性方法正在不断开发,以改善患者护理.
- 爱德华兹的Acumen IQ传感器和HemoSphere高级监测平台 (HS-IQ) 是一种用于CO估计的新型系统.
研究的目的:
- 评估HS-IQ系统估计的心脏输出量 (CO) 与通过热稀释 (TD) 测量的狗之间的一致性.
- 为了评估HS-IQ CO测量的精度,无论是辐射和背侧甲动脉.
主要方法:
- 八只专门繁殖的比格尔犬被麻醉并用动脉导管进行仪器仪表.
- 用TD和HS-IQ系统测量了心脏输出,用于辐射和背甲动脉部位.
- 在normovolemic和hypovolemic状态下进行测量,使用Bland-Altman方法分析,可接受的LOA<30%.
主要成果:
- 总共分析了21个 (辐射) 和24个 (脊椎甲足) 配对的CO测量结果.
- 辐射动脉HS-IQ与TD的偏差为-0.09 L/min,共识范围很大 (-2.66 到 2.49 L/min).
- 背甲动脉HS-IQ与TD的偏差为-0.26L/分钟,具有广泛的协议边界 (-2.76至2.24L/分钟).
结论:
- 使用HS-IQ系统进行心脏输出估计在技术上是可行的,但结果超出了临床上可接受的范围.
- 广泛的共识限制表明,HS-IQ CO估计目前不适合临床使用.
- 在临床应用之前需要进一步验证和潜在的系统改进.
更多相关视频
07:51Standardized Hemorrhagic Shock Induction Guided by Cerebral Oximetry and Extended Hemodynamic Monitoring in Pigs
Published on: May 21, 2019
7.4K
06:10Author Spotlight: Assessment of Cardiac Output Calculation by Thermodilution in Pigs for Effective Perfusion Flow During EVLP
Published on: June 28, 2024
498
相关概念视频
Equipments Used To Measure Blood Pressure
3.4K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.4K
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
764
Clinical manifestationsPeripheral Arterial Disease (PAD) manifests through a range of symptoms, from the characteristic intermittent claudication to atypical presentations and severe complications in advanced stages. Intermittent claudication, a hallmark symptom of PAD, presents as exercise-induced muscle pain that typically resolves within minutes of rest. This pain is reproducible and stems from inadequate blood flow, leading to the accumulation of lactic acid produced during anaerobic...
764