一种便携式微等离子光学发射光谱仪器,具有在线消化功能,用于现场和生物样本中的敏感测定
Ji-Ying Cai1, Xiao Zhang1, Yu-Jia Wei1
1Research Center for Analytical Sciences, Department of Chemistry, College of Sciences, Northeastern University, Box 332, Shenyang 110819, China.
Analytical chemistry
|February 19, 2024
概括
一个新的便携式设备可以快速地在现场检测生物样本中的 (Pb),使用点放电微等离子体光学发射光谱学. 这种方法为环境健康风险评估提供了灵敏而准确的选.
科学领域:
- 分析化学 分析化学
- 环境科学 环境科学
- 频谱学是一种光谱学.
背景情况:
- 在生物样本中准确检测 (Pb) 对于评估污染对人类健康的风险至关重要.
- 传统的方法很庞大,需要大量的样本准备,不适合现场分析.
研究的目的:
- 开发一个便携式的点放电 (PD) 微等离子体光学发射光谱 (OES) 装置,用于在生物样本中进行敏感的现场测定.
- 整合在线消化功能,以简化样本处理.
主要方法:
- 设计了一个带有集成在线消化模块的便携式PD-microplasma-OES设备.
- 样品经过使用HNO3和H2O2的快速消化,然后通过化物生成 (HG) 进行转化.
- 通过在微中通过激发和排放检测到,使用专门的采样柱.
主要成果:
- 在线消化六个50毫克的样品在25分钟内达到97%的效率.
- 对Pb的检测极限为0.018毫克/公斤-1,精度为3.6%.
- 该设备是紧的 (2.5公斤),低功率 (24.3瓦),适合现场使用,经过认证的参考材料和真实样本的验证.
结论:
- 开发的便携式PD-microplasma-OES设备为在生物矩阵中对等有毒元素的现场分析提供了方便和高效的工具.
- 这项技术有助于在环境和人类健康研究中快速评估污染的风险.
相关概念视频
Atomic Emission Spectroscopy: Lab
162
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
162
Gas Chromatography: Types of Detectors-II
371
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
371
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
221
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
221
Atomic Emission Spectroscopy: Instrumentation
422
The instrumentation of atomic emission spectrometry (AES) involves various components, including atomization devices that convert samples into gas-phase atoms and ions. There are two main types of atomization devices: continuous and discrete atomizers. Continuous atomizers, like plasmas and flames, introduce samples in a constant stream, while discrete atomizers inject individual samples using syringes or autosamplers. The most common discrete atomizer is the electrothermal atomizer.
422
Atomic Absorption Spectroscopy: Lab
389
For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
389