相关实验视频
Updated: Jul 2, 2025

06:01
Fluorescence Lifetime Macro Imager for Biomedical Applications
Published on: April 7, 2023
749
多光谱单点氧发光剂量测量 (MSOLD) 用于以光介导的光动力学疗法
Weibing Yang1, Vivek Rastogi1, Hongjing Sun1
1Department of Radiation Oncology, University of Pennsylvania, Philadelphia, PA 19104.
Proceedings of SPIE--the International Society for Optical Engineering
|February 29, 2024
概括
这项研究优化了光动力学疗法 (PDT) 的多谱单片氧剂量测量 (MSOLD),通过提高信号与噪声比. 增强的MSOLD方法在PDT应用中显示出准确的单点氧检测的前景.
科学领域:
- 生物医学光学 生物医学光学
- 光动力学疗法 光动力学疗法
- 医学物理 医学物理
背景情况:
- 单片氧 ([1O2) 的直接检测对于II型光动力学治疗 (PDT) 剂量测量至关重要.
- 目前的多光谱单片氧剂量测量 (MSOLD) 面临着由于单片氧信号的弱小和短暂而带来的挑战,需要改进信号噪声比.
研究的目的:
- 为了优化MSOLD信号采集,提高灵敏度和最小可检测性.
- 在各种体外条件下对单点氧显式剂量计 (SOED) 进行改进的MSOLD方法的验证.
- 为了准备在小鼠瘤模型中对MSOLD设备进行体内测试.
主要方法:
- 使用光纤检测和PDT治疗光的直角布局.
- 使用632nm激光和InGaAs探测器进行信号收集.
- 在1200-1300纳米的光谱辐射测量,然后进行背景减去和高斯适配,用于单个氧气信号量化.
主要成果:
- 该研究的重点是优化MSOLD信号采集并验证其灵敏度和最小可检测性.
- 将MSOLD信号与基于体内测量的单片氧显式剂量计 (SOED) 结果进行比较.
- 成功开发和验证了一种改进的MSOLD技术,用于单点氧气检测.
结论:
- 优化的MSOLD方法为PDT中直接单点氧气检测提供了改进的信号噪声比.
- 这一进步对于在II型PDT中准确的剂量测量至关重要.
- 经过验证的MSOLD设备已经准备好在临床前PDT模型中进行体外测试.
关键词:
在InGaAs的基础上,在MSOLD中使用MSOLD.在PDT中使用PDT.多光谱发光剂量计 (MSOLD) 是一种多光谱发光剂量计.从内部拍摄的照片单一氧显式剂量计 (SOED) 是一种单一氧显式剂量计.更多相关视频
相关概念视频
Photoluminescence: Fluorescence and Phosphorescence
2.0K
Photoluminescence is a process where a molecule absorbs light energy and re-emits it in the form of light. This phenomenon occurs when a substance absorbs photons, promoting its electrons to higher energy level excited states, followed by a relaxation process in which the electrons return to their original ground state energy levels and emit light. Photoluminescence is widely observed in various materials, including semiconductors, and organic and inorganic compounds.
A pair of electrons in a...
A pair of electrons in a...
2.0K
Photoluminescence: Applications
393
Photoluminescence offers a wide range of applications due to its inherent sensitivity and selectivity. This technique allows for both direct and indirect analyses of the analyte. Direct quantitative analysis is possible when the analyte exhibits a favorable quantum yield for fluorescence or phosphorescence. However, an indirect analysis may be feasible if the analyte is not fluorescent or phosphorescent, or if the quantum yield is unfavorable. Indirect methods include reacting the analyte with...
393