视觉章的基本特征及其作为内镜医生的眼镜剂量计的性能
Kotaro Fukushima1, Kosuke Matsubara2, Nao Ichikawa1,3
1Department of Quantum Medical Technology, Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80, Kodatsuno, Kanazawa, Ishikawa, Japan.
概括
视觉章,一种眼镜镜片剂量计,在内镜师的幻影研究中表现出令人满意的性能. 它显示出良好的线性和统一性,尽管具有特定的角度依赖性,并且通常与OSL剂量计趋势保持一致.
科学领域:
- 医学物理 医学物理
- 辐射剂量计 辐射剂量计
- 职业健康 职业健康 职业健康
背景情况:
- 内镜使用者暴露于辐射,需要精确的眼镜镜片剂量测量.
- 视觉章是一种新型的眼镜镜片剂量计,旨在测量Hp(3),在镜片中3mm深度的剂量.
- 评估其性能对于确保医疗程序中的辐射安全至关重要.
研究的目的:
- 评估视觉章眼镜剂量计的基本特征.
- 评估视觉章在模拟内镜师使用幻影研究的职业环境中的性能.
- 为了比较视觉章的测量与已建立的剂量测量方法.
主要方法:
- 进行了幻影研究,以评估能量依赖 (27.9-38.7 keV),剂量线性 (高达21.4 mGy),批量统一性 (变化系数为3.47%) 和角度依赖 (0-90°).
- 视觉章和光学刺激发光剂量计 (OSL) 被附在内镜师幻影的放射保护眼镜上.
- 视觉章的hp(3) 测量结果与OSL剂量计 (nanoDot) 进行了比较,并在各种辐射条件下进行了评估.
主要成果:
- 视觉章表现出良好的线性 (R2=1.00) 和可接受的批量统一性.
- 能源依赖在Hp(3) 中增加了3.8%,从27.9 keV增加到38.7 keV.
- 观察到特定的角度依赖性,测量下降到75°,在90°增加,在某些角度显示35.5%-55.0%小于nanoDot.
结论:
- 视觉章显示了令人满意的基本剂量测量特性,用于内镜医生对眼镜的监测.
- 它的特定角度依赖,归因于它的形状,在实际应用过程中需要考虑.
- 该研究表明,视觉章和OSL剂量计在幻影条件下存在相似的趋势.
关键词:
相当于剂量相当的剂量.一个剂量计.内镜逆行性胆血管造机图 (cholangiopancreatography) 是一种内镜逆行式胆血管造机图 (cholangiopancreatography).眼镜眼镜眼镜眼镜眼镜眼镜职业暴露 职业暴露 职业暴露更多相关视频
相关概念视频
Spectrophotometry: Introduction
8.5K
Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
8.5K
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
7.8K
Ultraviolet–visible (UV–visible or UV–Vis) spectroscopy is an analytical technique that investigates the interaction between matter and UV–Vis light within the electromagnetic spectrum. This method is widely used for its versatility, simplicity, and relatively quick data acquisition, making it valuable for both qualitative and quantitative analysis. When UV–Vis radiation passes through a material, molecules absorb light depending on the energy required for...
7.8K
UV–Vis Spectrometers
4.1K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
4.1K
Imaging Biological Samples with Optical Microscopy
9.2K
Optical microscopy uses optic principles to provide detailed images of samples. Antonie van Leeuwenhoek designed the first compound optical microscope in the 17th century to visualize blood cells, bacteria, and yeast cells. In 1830, Joseph Jackson Lister created an essentially modern light microscope. The 20th century saw the development of microscopes with enhanced magnification and resolution.
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
9.2K
Focusing of Light in the Eye
6.3K
Light rays enter the eye through the cornea, a transparent dome-shaped tissue that is the eye's outermost layer. The cornea bends or refracts, light rays traveling to the pupil. The shape of the cornea determines how much of the light is bent and whether the image will be focused correctly on the retina at the back of the eye. Once the light has passed through both refraction layers, it converges into a single focal point onto a small area. This is where photoreceptors start transforming...
6.3K
Ophthalmic Drug Delivery Systems
285
Ophthalmic drug delivery faces major limitations due to poor absorption across the corneal membrane. This process is primarily driven by diffusion and is influenced by two main factors: the physicochemical properties of the drug and tear drainage. Most ophthalmic drugs, such as pilocarpine, epinephrine, atropine, and local anesthetics, are weak bases. They are typically formulated at an acidic pH to enhance chemical stability. However, this leads to high ionization, reducing their ability to...
285