Optics letters
|April 15, 2024
PubMed
概括

研究人员使用环状相创建了具有非马科维特征的非衍射斑点. 这些光学场保持稳定的强度配置文件,并提供对统计属性的控制,促进对非马科夫过程的理解.

相关概念视频

¹H NMR: Interpreting Distorted and Overlapping Signals01:02

¹H NMR: Interpreting Distorted and Overlapping Signals

Spin systems where the difference in chemical shifts of the coupled nuclei is greater than ten times J are called first-order spin systems. These nuclei are weakly coupled, and their chemical shifts and coupling constant can generally be estimated from the well-separated signals in the spectrum.
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
1.0K
Time and frequency -Domain Interpretation of Phase-lead Control01:24

Time and frequency -Domain Interpretation of Phase-lead Control

Phase-lead controllers are commonly used in various control systems to enhance response speed and stability. Adjusting the brightness on a television screen offers a practical example of phase-lead control. When contrast is enhanced, a phase-lead controller is employed. Mathematically, phase-lead control is identified when the first parameter is smaller than the second.
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
82
Phase Contrast and Differential Interference Contrast Microscopy01:26

Phase Contrast and Differential Interference Contrast Microscopy

Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
8.0K
NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
692
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)01:20

¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)

When proton-coupled carbon-13 spectra are simplified by a broadband proton decoupling technique, structural information about the coupled protons is lost. Distortionless enhancement by polarization transfer (DEPT) is a technique that provides information on the number of hydrogens attached to each carbon in a molecule. While the DEPT experiment utilizes complex pulse sequences, the pulse delay and flip angle are specifically manipulated. The resulting signals have different phases depending on...
1.0K