在急诊室和重症监护室进行非侵入性持续血压监测的直接评估
Journal of emergency nursing
|April 19, 2024
概括
非侵入性持续血压监测对医院来说是有前途的,但VitalStream设备需要进一步开发. 目前的准确性不符合临床标准,用于重症监护决策.
科学领域:
- 生物医学工程 生物医学工程
- 临床监测 临床监测
- 医疗器械 医疗器械
背景情况:
- 非侵入性的连续血压监测有可能在医院更早地检测患者的病情恶化.
- 它可以赋予临床医生权力,并减少与侵入性监测方法相关的风险.
- 新兴技术必须与已建立的临床措施对准确性进行验证.
研究的目的:
- 评估护士的可用性和愿意采用VitalStream非侵入性连续血压计.
- 为了比较VitalStream的血压读数与"黄金标准"方法 (手臂袖,动脉线).
主要方法:
- 评估了VitalStream设备的可用性和实施意愿.
- 在VitalStream和黄金标准措施之间比较了静脉缩压,静脉缩压和平均动脉压.
- 数据是在紧急和重症监护病房设置中收集的.
主要成果:
- 两个机构的护士都接受了类似的设备,急救护士的认可率更高.
- VitalStream的测量结果没有满足关于可接受的血压测量差异的既定共识标准.
- 布兰德-阿尔特曼分析显示偏差标准偏差超过8mmHg,并<85%的对差在10mmHg范围内.
结论:
- 目前的VitalStream设备需要额外的后处理,以便在紧急情况下可靠地做出医疗决策.
- 需要进一步精细化,以满足临床准确性标准,用于重症监护应用.
相关概念视频
Pulse rhythm
790
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
790
Special considerations while measuring blood pressure
722
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
722
Equipments Used To Measure Blood Pressure
965
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
965
Errors occurring during blood pressure monitoring
719
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
719
Assessing Blood pressure using a doppler ultrasound
1.3K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.3K
Assessment of blood pressure in brachial artery(two-step method)
704
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
704