Robert L Schrader1, Gordon A Anderson2, David H Russell1
1Department of Chemistry, Texas A&M University, College Station, Texas 77843, United States.
Journal of the American Society for Mass Spectrometry
|April 30, 2024
概括
这项研究评估了针对离子隔离的细分四极质量过器. 虽然在宽峰上实现了100%的效率,但在较窄的质量-电荷比上,性能显著下降.
科学领域:
- 分析化学 分析化学
- 质谱测量质量谱测量
背景情况:
- 四极质量过器对于质谱学中的离子选择至关重要.
- 优化离子隔离效率对于复杂混合物分析至关重要.
研究的目的:
- 为了评估一个带有矩形波形的细分四极质量过器的性能.
- 为了确定不同峰值宽度的离子隔离效率.
主要方法:
- 使用的是自制的四极轨道平台.
- 矩形和电容合的矩形波形应用于预过器.
- 用SIMION模拟来进行峰值形状分析.
主要成果:
- 对于50m/z的峰值宽度,实现了100%的隔离效率.
- 在150V0-p的5m/z峰值宽度下,效率下降到20%左右.
- 观察到峰值结构是由于一个小的出口口和有限的射频电压重定位.
结论:
- 分段四极管减轻了边缘场的失焦,但提供了有限的重定焦.
- 在隔离效率方面可以实现的收益受到射频电压和系统设计的限制.
- 为了在质谱学中实现高分辨率离子隔离,需要进一步优化.
相关概念视频
Mass Analyzers: Common Types
600
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
600
Mass Analyzers: Overview
661
The mass analyzer is a crucial component of the mass spectrometer. In the ionization chamber, the vaporized sample is bombarded with a high-energy electron beam to generate a radical cation and further fragment into neutral molecules, radicals, and cations. A series of negatively charged accelerator plates accelerate the cations into the mass analyzer. The mass analyzer separates ions according to their mass-to-charge (m/z) ratios and then directs them to the detector. The common types of mass...
661
Mass Spectrometers
5.4K
This lesson details the instrumentation of a mass spectrometer—a physical instrument to perform mass spectrometry on analyte molecules and record the characteristic mass spectra. This is achieved via three chief functions:
5.4K
Tandem Mass Spectrometry
984
Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and signal-to-noise ratio for the analyte. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
984
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
725
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
725
High-Resolution Mass Spectrometry (HRMS)
1.4K
The resolution of a mass spectrometer depends on the efficiency of separating ions with different ion masses. The mass of an atom is approximated to the sum of the masses of protons and neutrons inside, considering the masses of protons and neutrons as equal. However, the masses of the proton (1.6726 × 10−24 g) and neutron (1.6749 × 10−24 g) are not truly equal. There is a minor error in the expression of atomic masses relative to the simplest atom of hydrogen. For...
1.4K