与胺分析相关的吸附性损失带来的分析前挑战
Katie A Edwards1,2, Eileen A Randall3, Patricia C Wolfe4
1Department of Pharmaceutical Sciences, Binghamton University, Binghamton, NY, 13902, USA. kedwards@binghamton.edu.
Scientific reports
|May 4, 2024
概括
玻璃器皿和过器在样本处理过程中会导致大量的胺 (维生素B1) 损失. 建议使用聚烯和尼龙等聚合物材料作为过器和瓶子,以防止不准确的维生素B1测量.
科学领域:
- 生物化学 生化学
- 分析化学 分析化学
- 环境科学 环境科学
背景情况:
- 胺 (维生素B1) 对代谢途径至关重要,以其二酸盐形式作为辅因子.
- 临床和环境样本中低度的胺需要经过仔细的分析前处理.
- 在实验室中普遍使用玻璃材料可能会影响胺回收.
研究的目的:
- 调查分析前加工材料对胺回收的影响.
- 为了识别导致胺损失的特定材料.
- 为准确的胺分析推替代材料.
主要方法:
- 评估在玻璃瓶中储存后从标准中恢复胺与聚烯瓶子.
- 在通过玻璃纤维过器后评估胺酸度.
- 分析不同储存容器中胺及其化衍生物的分布.
主要成果:
- 在玻璃纤维过器和玻璃瓶中观察到显著的胺损失.
- 与聚烯相比,在玻璃瓶中储存导致胺峰值面积减少了54%.
- 聚合物过器 (尼龙,纤维素) 和容器 (聚碳酸盐,聚烯) 显示了最小的胺损失.
结论:
- 玻璃材料 (过器,瓶子) 在样本制备过程中显著降低了胺度.
- 建议在过器和样品储存中使用聚合物材料,以确保精确的胺量化.
- 精心选择预分析材料对于可靠的胺分析至关重要,特别是在低度下.
相关概念视频
Atomic Absorption Spectroscopy: Lab
349
For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
349
Voltammetry: Stripping Methods
208
Anodic Stripping Voltammetry (ASV), Cathodic Stripping Voltammetry (CSV), and Adsorptive Stripping Voltammetry (AdSV) are electrochemical techniques used to determine trace amounts of analytes in solution. These methods involve applying a potential to an electrode and measuring the resulting current.
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
208
Atomic Absorption Spectroscopy: Interference
745
Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
745