概括
从基线BIMS数字握力动力表和Jamar+的手柄强度测试结果是可互换的. 这项研究发现这两种设备之间有很好的一致性和标准有效性,这表明可靠的身体状况评估的测量误差很低.
科学领域:
- 生物医学工程 生物医学工程
- 运动学 运动学
- 老年学是一门学科.
背景情况:
- 手握强度 (HGS) 是整体身体健康和功能的关键指标.
- 标准化的HGS测量对于准确的健康评估至关重要.
- 有各种各样的动力计,需要验证可互换使用.
研究的目的:
- 评估基准BIMS数字抓地力表与Jamar电子动力表 (Jamar+) 之间的仪器间协议和标准有效性.
- 确定这两种设备的HGS测量是否可以在临床和研究环境中互换使用.
主要方法:
- 70名参与者 (年龄在23-88岁之间) 用随机顺序对每个动力计进行了三次HGS试验.
- 计算了类内相关系数 (ICC),标准测量误差 (SEM) 和最小可检测变化 (MDC).
- 使用了新的,制造商校准的动力计 (基准BIMS和Jamar+).
主要成果:
- 在两个动力计之间观察到0.98的优秀整体ICC.
- 对于最强的尝试 (0.68公斤) 和三次尝试的平均值 (0.37公斤) 发现了最小的平均差异.
- 与Jamar+相比,SEM和MDC值表明基线BIMS设备的测量误差较低.
结论:
- 与Jamar+相比,基线BIMS数字接动力表显示出高度一致性和标准有效性.
- 结果表明,从两个动力计获得的测量值可以互换使用.
- 这有助于在不同设备之间进行一致和可靠的HGS评估.
相关概念视频
Equipments Used To Measure Blood Pressure
964
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
964
Assessment of blood pressure in brachial artery(two-step method)
698
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
698