在切片和厚度测量过程中持续冷却样本,有助于提高病理样本图像分析的准确性
Masanao Saio1, Raia Kushibiki2, Yuki Kanehira2
1Laboratory of Histopathology and Cytopathology, Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences, Gunma, Japan. saio@gunma-u.ac.jp.
Histology and histopathology
|May 6, 2024
概括
连续冷却装置 (CCD) 提高了组织样本的断面厚度 (ST) 均性. 一致的ST对于在组织病理学中进行准确的定量分析至关重要,确保可靠的染色结果.
科学领域:
- 组织病理学 组织病理学
- 生物医学成像技术 生物医学成像技术
- 材料科学 材料科学 材料科学
背景情况:
- 在组织病理学中,准确的定量分析依赖于一致的样本截面厚度 (ST).
- 在ST的变化可以显著影响的正染色的百分比面积,影响诊断和研究结果.
研究的目的:
- 为了研究断面厚度 (ST) 与正色彩的百分比面积之间的关系.
- 评估冷却方法 (冰包与连续冷却装置) 对ST统一性的影响.
主要方法:
- 从仅含和FFPE块使用冰袋 (IP) 或连续冷却装置 (CCD) 准备了截面.
- 用薄膜厚度计测量了截面厚度.
- 染色是使用欧或氨酸蓝色进行的.
- 计算机辅助图像分析量化了阳性染色的百分比面积及其与ST的相关性.
主要成果:
- 与冰箱 (IP) 相比,连续冷却装置 (CCD) 产生了明显更薄,更均的截面.
- 在ST和欧和氨酸蓝的阳性染色的百分比面积之间发现了显著的相关性.
- 与具有变异性ST的部分相比,具有均ST的部分显示出明显不同的染色区域.
结论:
- 建议使用连续冷却装置 (CCD) 进行样本切割,以实现统一的ST.
- 在染色前测量ST对于准确的定量组织病理学分析至关重要.
- 组织学检查应使用具有统一ST的标本,以获得可靠的结果.
相关概念视频
Fixation and Sectioning
4.3K
Two basic types of preparation are used to visualize specimens with a light microscope: wet mounts and fixed specimens.
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
4.3K
Phase Contrast and Differential Interference Contrast Microscopy
7.9K
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
7.9K