一个回顾性数据库分析,以调查静脉血液中的电解质是否相当于动脉血液中的水平
R N Devaki1, Prajna Kasargod1, An Roopa Urs2
1Department of Biochemistry, Chamarajanagar Institute of Medical Sciences, Chamarajanagar, Karnataka, India.
概括
动脉血液气体 (ABG) 分析提供了快速的电解质测量. 然而,这项研究发现,在重症患者中,ABG电解质值不能可靠地替代常规血清电解质分析.
科学领域:
- 临床化学 临床化学
- 临界护理医学 临界护理医学
- 实验室诊断 实验室诊断 实验室诊断
背景情况:
- 动脉血液气体 (ABG) 分析仪在血液气体分析的同时提供快速的电解质测量.
- 对ABG分析的周转时间比传统的血清电解质检测要快得多.
研究的目的:
- 评估通过ABG分析测量动脉血液中的电解质是否可以取代常规血清电解质测量.
- 为了比较从动脉和静脉血液样本中获得的电解质值的准确性和一致性.
主要方法:
- 一项回顾性横截面研究分析了200名患者的数据.
- 从同时动脉 (ABG) 和静脉 (血清) 血液样本中的电解质水平进行了比较.
- 统计分析包括一致性相关系数和布兰德-阿尔特曼图.
主要成果:
- 平均,和含量显示动脉和静脉样本之间的差异.
- 通过ABG分析的电解质值和使用血清样本的常规自分析仪之间没有发现显著的一致性.
- 布兰德-阿尔特曼分析表明缺乏一致性.
结论:
- 动脉血液气体分析仪的电解质测量与使用血清样本的传统自动分析仪的结果不可互换.
- 常规血清电解质分析仍然是标准的护理,不能被ABG电解质测量所取代.
相关概念视频
Electrogravimetric Analysis: Overview
220
Electrogravimetric analysis measures the weight of an analyte deposited electrolytically onto a suitable working electrode. This method involves applying a potential to a pre-weighed electrode submerged in a solution, which results in the desired substance being deposited through reduction at the cathode or oxidation at the anode. The electrode's weight is recorded after deposition, and the difference in weight gives the analyte's weight in the solution.
To test the completeness of the...
To test the completeness of the...
220
Blood Studies I: ABG and VBG
495
Blood studies are critical in the medical field, enabling healthcare professionals to assess a patient's health status accurately. This page will focus on two significant blood studies: Arterial Blood Gas (ABG) and Venous Blood Gas (VBG).
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
495
EDTA: Indirect and Alkalimetric Titration
901
Unlike direct titration, back-titration, and displacement titration, indirect titration is an EDTA titration method for quantifying anions. In the indirect titration method, anions are precipitated as their insoluble salts with excess metal ions. The filtrate containing the excess metal ions is directly titrated with standard EDTA until the endpoint is achieved. Another approach involves extracting the metal ion and back-titrating with standard EDTA to obtain the endpoint. In this way, the...
901
Roles of Electrolytes: Chloride and Bicarbonate
209
Chloride ions contribute to the osmotic pressure gradient distinguishing the intracellular fluid (ICF) from the extracellular fluid (ECF). They counterbalance positively charged ions in the ECF and ensure its electrochemical stability. The renal system's process of chloride absorption and release generally mirrors that of sodium ions.
Conditions such as hypochloremia can arise from insufficient chloride reabsorption by the kidneys, often compounded by extended bouts of diarrhea, vomiting,...
Conditions such as hypochloremia can arise from insufficient chloride reabsorption by the kidneys, often compounded by extended bouts of diarrhea, vomiting,...
209
Effects of EDTA on End-Point Detection Methods
272
Different methods, such as visual observance of metal-ion indicators, spectroscopic techniques, and potentiometric methods, can determine the endpoint of an EDTA titration.
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
272
Electrodes: Overview
1.7K
Electrochemical measurements are conducted in an electrochemical cell composed of various components that control and measure the current and potential. One fundamental component is electrodes, conductive materials that enable electron transfer reactions at their surfaces.
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
1.7K