用MR弹性图进行刚度测量的精度和测试-重复测试可重复性:一个多中心幻影研究
Efe Ozkaya1, Paul Kennedy1, Jun Chen1
1From the BioMedical Engineering and Imaging Institute (E.O., P.K., O.B., B.T.) and Departments of Diagnostic, Molecular and Interventional Radiology (E.O., P.K., O.B., B.T.) Environmental Medicine and Public Health (J.T.D.), Icahn School of Medicine at Mount Sinai, 1470 Madison Ave, New York, NY 10029; Department of Radiology, Mayo Clinic, Rochester, Minn (J.C., P.J.R., S.K.V., R.L.E.); Department of Radiology, Nanjing University Medical School Affiliated Drum Tower Hospital, Nanjing, China (J.C.); Department of Radiology, Cincinnati Children's Hospital Medical Center, University of Cincinnati College of Medicine, Cincinnati, Ohio (J.R.D., J.A.T.); Joint Department of Medical Imaging, University Health Network, Mount Sinai Hospital, and Women's College Hospital, University of Toronto, Toronto, Canada (K.S.J.); Department of Radiology and Biomedical Imaging, University of California, San Francisco, Calif (M.A.O.); and Department of Radiology, Zuckerberg San Francisco General Hospital, San Francisco, Calif (M.A.O.).
在多中心幻影研究中,MR弹性图 (MRE) 证明了肝硬度测量的高精度和可重复性. 标准化的协议确保可靠的非侵入性肝纤维化在不同扫描仪的分期.
科学领域:
- 医疗成像医学成像
- 生物物理学的生物物理.
- 量化MRI是指数量化的MRI.
背景情况:
- 核磁共振弹性图 (MRE) 是一种用于非侵入性肝纤维化病阶段的验证技术.
- 关于在多中心环境中MRE刚度测量精度和可重复性的数据有限.
研究的目的:
- 在多个机构中评估MRE刚度测量的精度和测试-重复测试可重复性.
- 通过使用标准化幻象和获取协议来评估MRE性能.
主要方法:
- 在五个中心使用10个临床MRE系统 (1.5-T和3-T) 扫描了三种具有不同硬度的聚乙烯凝幻影.
- 采用了标准化的二维 (2D) 梯度召唤回声 (GRE) 和二维旋回声 (SE) 回声平面成像 (EPI) 协议.
- 计算了刚度精度 (变量系数,CV) 和可重复性 (可重复性系数,RC).
主要成果:
- 总体平均精度 (CV) 为5.8%; 2D GRE显示4.5%的CV,2D SE EPI显示7.8%的CV.
- 总体平均测试重复测试可重复性 (RC) 为5.8%; 2D GRE实现了4.9%的RC,2D SE EPI实现了7.0%的RC.
- 在各中心观察到卓越的体外精度和可重复性.
结论:
- 在多中心环境中,MR弹性显微镜表现出卓越的体外精度和相同会话测试-重复测试可重复性.
- 标准化的MRE协议,硬件和重建算法是实现一致硬度测量的关键.
- 这些发现支持MRE在多中心肝纤维化阶段测定中的可靠性.