概括
本研究引入了一种用于调整超声波声学输出的自动化方法,通过针对特定的信号噪声比 (SNR) 值来确保安全. 这种适应性方法旨在在胎儿成像期间使用一致的低水平机械指数 (MI).
科学领域:
- 医疗成像医学成像
- 超声波技术 超声波技术 超声波技术
- 生物医学工程 生物医学工程
背景情况:
- 超声波的声学输出调节不一致地被超声波仪所应用,尽管有安全建议.
- 目前的做法并不总是遵守超声波安全的最低合理可实现 (ALARA) 原则.
研究的目的:
- 开发和评估一种新的自动化方法,用于超声波中的自适应声学输出调整.
- 根据时间信号噪声比率 (SNR) 指导机械指数 (MI) 输出,以确保安全和图像质量.
主要方法:
- 开发了一种自适应算法,以根据28dB的目标时间信号噪声比 (SNR) 调整机械指数 (MI).
- 该方法在临床超声波系统上实施,并在胎儿腹部成像期间在11名志愿者身上体内测试.
- 在静止和动态扫描条件下评估性能,评估所需的MI和SNR维护.
主要成果:
- 需要28dBSNR的平均MI范围从0.17到0.26,随着成像频率的提高而增加.
- 在90%的评估中,固定视图和71%的动态扫描保持了目标SNR水平.
- 该系统成功地适应MI输出以应对不断变化的成像条件.
结论:
- 自动,时间SNR引导的自适应声输出调整对于体内胎儿成像是可行的和有效的.
- 这种方法通过在低机械指数 (MI) 水平上保持最佳图像质量来促进一致遵守ALARA原则.
- 这些发现支持采用自适应声输出策略,以提高超声波安全性和图像一致性.
相关概念视频
Design Example: Vintage Mixing Console
A sound engineer at a music company recently encountered a problem. The output from their newly acquired studio's vintage mixing console was too low for the requirements of modern recording equipment. To rectify this situation, the engineer decided to design an audio pre-amplifier using an operational amplifier (op-amp) to boost the signal level.
The specifications for the pre-amplifier were clear. It needed to amplify the audio signal by a factor of 10, have an input impedance above 10...
The specifications for the pre-amplifier were clear. It needed to amplify the audio signal by a factor of 10, have an input impedance above 10...
Properties of Fourier Transform I
The application of Fourier Transform properties in radio broadcasting is multifaceted, enabling significant advancements in the way signals are transmitted and received. Key areas where these properties are utilized include simultaneous multi-channel transmission, audio clip speed adjustments, live broadcast delays for different time zones, audio frequency adjustments, and signal demodulation.
In radio broadcasting, multiple audio signals often need to be transmitted simultaneously. The Fourier...
In radio broadcasting, multiple audio signals often need to be transmitted simultaneously. The Fourier...
Downsampling
When considering a sampled sequence with zero values between sampling instants, one can replace it by taking every N-th value of the sequence. At these integer multiples of N, the original and sampled sequences coincide. This process, known as decimation, involves extracting every N-th sample from a sequence, thereby creating a more efficient sequence.
The Fourier transform of the decimated sequence reveals a combination of scaled and shifted versions of the original spectrum. This...
The Fourier transform of the decimated sequence reveals a combination of scaled and shifted versions of the original spectrum. This...
Upsampling
Managing signal sampling rates is essential in digital signal processing to maintain signal integrity. A decimated signal, characterized by a reduced frequency range due to its lower sampling rate, can be upsampled by inserting zeros between each sample. This upsampling process expands the original spectrum and introduces repeated spectral replicas at intervals dictated by the new Nyquist frequency. To refine this zero-inserted sequence, it is passed through a lowpass filter with a cutoff...