怀孕期间的宫弹性的量化基于阴道超声成像和压力测量
IEEE transactions on bio-medical engineering
|May 21, 2024
概括
一个新的定量宫弹性学系统通过结合应力传感器,准确地测量宫弹性. 这个系统显示出怀孕期间宫显著软化,每周下降,并在三个月内变化.
科学领域:
- 生物医学工程 生物医学工程
- 医疗成像医学成像
- 产科 产科 产科 产科 产科
背景情况:
- 现有的宫弹性学方法,如应变和剪切波弹性学,在比较跨会话的弹性和处理组织不均性方面存在局限性.
- 对宫弹性的定量评估对于了解妊娠进展和识别早产等风险至关重要.
研究的目的:
- 开发和验证一种新的定量宫弹性体系统,克服现有方法的局限性.
- 为了使宫弹性的准确,稳固和可重复测量,用于纵向和检查员之间的比较.
主要方法:
- 将压力传感器集成到透阴超声系统中,以测量探头表面应力.
- 从B模式图像记录宫变形并使用自定义算法量化应变.
- 通过应力-应变线性回归估计宫阳模量.
主要成果:
- 幻影实验证明了开发系统的高精度,稳定性,可重复性和可重复性.
- 在孕妇参与者中,每周宫阳模量 (3.95%) 显著下降.
- 平均模值从第一季度的13.07 kPa下降到第三季度的4.40 kPa.
结论:
- 开发的定量宫弹性学系统是准确的,坚固的,安全的,并促进可靠的弹性测量.
- 该系统对不同超声波机器的适应性支持对怀孕期间宫软化模式的大规模研究.
- 这项技术可以为诸如早产等情况提供宝贵的见解.
相关概念视频
Stress-Strain Diagram
653
A stress-strain diagram is a crucial tool that graphically displays a material's mechanical characteristics. This diagram is derived from a tensile test performed on a carefully prepared cylindrical specimen. The specimen has two gauge marks inscribed on its central part, and the distance between these marks is known as the gauge length. The cylindrical specimen is placed in a testing machine, which applies an increasing centric load. As this load grows, so does the gauge length. This...
653
True Stress and True Strain
297
Engineering stress is calculated as the load divided by the original, undeformed cross-sectional area. It approximates a material under load. This approximation is especially relevant post-yield in ductile materials. Though engineering stress-strain diagrams are often used for their convenience and accessibility, they can sometimes fall short in accuracy, particularly when dealing with large strain values.
In contrast, true stress offers a more precise portrayal. It is computed by dividing the...
In contrast, true stress offers a more precise portrayal. It is computed by dividing the...
297
Elasticity in Concrete
92
Upon subjecting concrete to moderate or high uniaxial compressive or tensile stresses, the strain response is non-linear relative to the stress applied. As the stress is removed, the resulting stress-strain curve deviates from the original path traced during loading, creating a hysteresis loop, indicative of the concrete's non-linear and non-elastic properties. Typically, a material's modulus of elasticity, which is a measure of the material's stiffness, is inferred from the linear...
92
Measurements of Strain
731
Strain quantifies the deformation of a material under force, typically measured as normal strain, which represents the change in length when compared with the original length. Electrical strain gauges are used for enhanced accuracy. These devices consist of a conductive wire mounted on a paper backing that adheres to the material's surface. These gauges operate on the piezoresistive effect, where the wire's electrical resistance changes in response to mechanical deformation. The strain...
731