为医疗和美学应用设计的火花等离子装置的开发和表征
Mohammad Reza Lotfi1, Mohammadreza Khani1, Abootaleb Moradi1
1Laser and Plasma Research Institute, Shahid Beheshti University, G.C., Evin, 1983963113, Tehran, Iran.
Heliyon
|July 12, 2024
概括
这项研究开发了一种安全的冷大气火花等离子体装置,用于医疗用途. 电气和辐射安全测试证实其适用于治疗的适用性,并确定了有益的活性物种.
科学领域:
- 等离子体物理学的物理学
- 生物医学工程 生物医学工程
- 医疗器械开发 医疗器械开发
背景情况:
- 冷大气等离子体 (CAP) 设备在医疗应用方面显示出前景.
- 开发安全有效的CAP工具对于临床翻译至关重要.
- 现有的设备需要对医疗和美学用途进行彻底的表征.
研究的目的:
- 为安全的医疗和美学应用开发和描述冷大气火花等离子体装置.
- 评估火花等离子体的工程方面和免疫测量,以评估其治疗潜力.
- 确保设备符合用户健康的安全标准.
主要方法:
- 设备方案设计和阻断信号测试.
- 手持件的设计是为了提高安全性.
- 测试电气性能 (电流水平).
- 辐射功率评估 (紫外线光谱).
- 用于物种分析的光学发射光谱 (NOx,OH).
主要成果:
- 设备的电流低于安全值.
- 辐射功率达到危险水平的时间远远长于典型的治疗时间.
- 没有发现紫外线光谱辐射带来的风险.
- 光学发射光谱证实了NOx和OH等反应性物种的存在.
结论:
- 开发的冷大气火花等离子装置是安全的医疗和美学应用.
- 该设备的设计和特征性质符合安全要求.
- 鉴定的反应性物种 (NOx,OH) 表明潜在的治疗益处.
相关概念视频
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
573
Inductively coupled plasma (ICP) is the most widely used plasma source in atomic emission spectroscopy (AES), also known as Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES). The ICP source, or torch, consists of three concentric quartz tubes with argon gas flowing through them. A spark from a Tesla coil initiates the ionization of argon, generating a high-temperature plasma.
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
573
Atomic Emission Spectroscopy: Overview
2.0K
Atomic emission spectroscopy (AES) is an analytical technique used to determine the elemental composition of a sample by analyzing the light emitted from excited atoms. In AES, atoms in a sample are excited to higher energy levels by thermal energy from high-temperature sources, such as plasma, arcs, or sparks. When these excited atoms return to lower energy states, they emit light at specific wavelengths characteristic of each element. The resulting atomic emission spectrum, which consists of...
2.0K
Atomic Emission Spectroscopy: Lab
155
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
155
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
208
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
208