在磁场中使用不同半径的Farmer型电离室进行碳离子束剂量测量:测量和模拟
S Surla1,2,3, M Marot1,3,4, L Burigo1,3
1German Cancer Research Center DKFZ, Medical Physics in Radiation Oncology, Heidelberg, Germany.
Physics in medicine and biology
|July 25, 2024
概括
高达1特斯拉的磁场对碳离子束中的电离室反应产生非线性影响,并按室体积计算. 确定了新的磁场校正因子,用于农家型室,对于精确的辐射剂量测量至关重要.
科学领域:
- 医学物理 医学物理
- 辐射剂量计 辐射剂量计
- 微粒疗法是一种微粒疗法.
背景情况:
- 精确的剂量测量在碳离子治疗中至关重要.
- 外部磁场可以影响辐射检测设备.
- 了解这些效应对于磁场中可靠的剂量测量至关重要.
研究的目的:
- 研究磁场对碳离子束中电离室反应的影响.
- 为了确定各种农民型电离室的磁场校正系数.
- 将实验测量与蒙特卡洛模拟进行比较.
主要方法:
- 在高达1T的磁场中,测量了6个Farmer类型腔室 (R1-R6) 的响应.
- 使用了一个实验电磁体和一个单能碳离子束 (390.75 MeV u-1).
- 执行蒙特卡洛模拟来计算局部剂量变化并验证实验数据.
主要成果:
- 室内对磁场的反应是非线性和体积依赖的,最大变化为0.30%至0.62%.
- 较低能量的光束显示,在室R3.3的0.2T和0.7T之间,响应系统下降 (0.2%).
- 计算的磁场校正系数在0.28% (R4) 到0.60% (R5) 之间.
结论:
- 这项研究提供了对电离室对高达1 T的磁场反应的第一个详细分析.
- 对于多个室内体积,磁场校正因子被成功计算出来.
- 这些发现对于在使用磁场的碳离子疗法中准确的剂量确定至关重要.
相关概念视频
Biological Effects of Radiation
15.4K
All radioactive nuclides emit high-energy particles or electromagnetic waves. When this radiation encounters living cells, it can cause heating, break chemical bonds, or ionize molecules. The most serious biological damage results when these radioactive emissions fragment or ionize molecules. For example, α and β particles emitted from nuclear decay reactions possess much higher energies than ordinary chemical bond energies. When these particles strike and penetrate matter, they...
15.4K
Mass Analyzers: Common Types
586
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
586
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
848
An applied magnetic field causes the electrons present in the molecule to circulate, setting up a local diamagnetic current within the molecule. The local diamagnetic current arising from circulating sigma-bonding electrons induces a magnetic field, Blocal that opposes the applied magnetic field, B0. The effective magnetic field experienced by these nuclei is given by the difference between the applied and local magnetic fields in a phenomenon called local diamagnetic shielding. Essentially,...
848
Mass Analyzers: Overview
627
The mass analyzer is a crucial component of the mass spectrometer. In the ionization chamber, the vaporized sample is bombarded with a high-energy electron beam to generate a radical cation and further fragment into neutral molecules, radicals, and cations. A series of negatively charged accelerator plates accelerate the cations into the mass analyzer. The mass analyzer separates ions according to their mass-to-charge (m/z) ratios and then directs them to the detector. The common types of mass...
627
Chemical Ionization (CI) Mass Spectrometry
722
The molecular ion peak of a molecule in the mass spectrum provides vital information for molecular identification. However, conventional electron impact ionization can lead to the rapid dissociation of some molecular ions before they reach the detector. A milder ionization method is required to increase the lifetime of such ionized analyte molecules. Chemical ionization (CI) is a gas-phase protonation reaction useful for mass-analyzing analyte molecules that are easily protonated to yield the...
722
Mass Spectrometers
5.3K
This lesson details the instrumentation of a mass spectrometer—a physical instrument to perform mass spectrometry on analyte molecules and record the characteristic mass spectra. This is achieved via three chief functions:
5.3K