增强修改的聚氨酸电压测量传感器用于定量检测唾液素
Renad S El-Kamel1, Amany M Fekry1
1Cairo University, Faculty of Science, Department of Chemistry, Giza, Egypt.
International journal of biological macromolecules
|July 27, 2024
概括
一种新型的电化学传感器使用碳糊电极 (pTyr/CPE) 上的聚合型氨酸,可以检测唾液氨酸. 这为诊断胃食道逆流性疾病 (GERD) 提供了一种非侵入性方法.
科学领域:
- 电化学 电化学 电化学
- 生物传感器是一种生物传感器.
- 分析化学 分析化学
背景情况:
- 素是一种关键的消化酶,对于检测胃食道逆流性疾病 (GERD) 至关重要.
- 当前的诊断方法可能是侵入性的或不舒服的,需要更简单的替代方案.
- 唾液提供一种非侵入性和可访问的生物流体,用于疾病生物标志物检测.
研究的目的:
- 开发一个负担得起和敏感的电化学传感器用于唾液素检测.
- 使用传感器进行GERD的初步诊断.
- 为了研究聚合型氨酸在碳糊电极 (pTyr/CPE) 上作为氨酸传感器的潜力.
主要方法:
- 在碳糊电极 (pTyr/CPE) 上通过电聚合合成聚合型氨酸薄膜的合成.
- 使用循环电压测量 (CV),微分脉冲电压测量 (DPVs),时频测量 (CA) 和电化学阻抗光谱 (EIS) 的电化学表征.
- 使用扫描电子显微镜 (SEM) 进行表面形态分析.
主要成果:
- 在pTyr/CPE中,在pH值为2.0的酸盐缓冲器中,表现出对素的敏感电压测量检测.
- 传感器显示了两个线性范围 (1-20 ng/mL和20-100 ng/mL),检测极限较低 (0.5 ng/mL和0.09 ng/mL).
- 观察到高选择性,稳定性和1.5秒的快速响应时间.
结论:
- pTyr/CPE传感器是通过唾液素检测初步诊断GERD的一个有希望的工具.
- 该传感器在从喉回流 (LPR) 患者的唾液样本中量化素水平方面表现出可靠的性能.
- 这种非侵入性的电化学方法为GERD诊断提供了一个可行的替代方案.
更多相关视频
08:37Sampling Human Indigenous Saliva Peptidome Using a Lollipop-Like Ultrafiltration Probe: Simplify and Enhance Peptide Detection for Clinical Mass Spectrometry
Published on: August 7, 2012
11.4K
05:33Measuring Biophysical and Psychological Stress Levels Following Visitation to Three Locations with Differing Levels of Nature
Published on: June 19, 2019
8.5K
相关概念视频
Potentiometry: Membrane Electrodes
2.3K
Membrane electrodes, also known as p-ion electrodes, use membranes that selectively interact with free analyte ions, generating a potential difference across the membrane. The resulting membrane potential, known as the asymmetry potential, is not zero even when analyte concentrations on both sides of the membrane are equal. The membrane's response is typically not selective to a single analyte but proportional to the concentration of all ions in the sample solution capable of interacting at...
2.3K
Amperometry: Overview
2.2K
Amperometry is a technique commonly used to measure the concentration of specific analytes in a solution by monitoring the electric current generated during an electrochemical reaction. It involves applying a constant potential between a working electrode and a reference electrode to measure the resulting current, which is proportional to the concentration of the analyte. The Clark oxygen electrode operates based on this principle of amperometry. It consists of a cathode and an anode enclosed...
2.2K
Capillary Electrophoresis: Instrumentation
1.9K
Capillary electrophoresis instrumentation typically consists of several key components. A high-voltage power supply generates the electric field necessary for the separation by connecting to an anode (the positively charged electrode) and a cathode (the negatively charged electrode) located in buffer reservoirs at each end of the capillary tube. The system includes a sample vial, a fused silica capillary tube coated with polyimide for mechanical strength through which the sample components...
1.9K