使用用于辐射治疗的充电发生器评估电表校准系数的每月变化
Motohiro Kawashima1, Maria Varnava2, Shuichi Ozawa3,4,5
1Gunma University Heavy Ion Medical Center, 3-39-22 Showa-Machi, Maebashi, Gunma, 371-8511, Japan. mkawashi@gunma-u.ac.jp.
Radiological physics and technology
|August 1, 2024
概括
对于大多数设备,电表校准系数 (kelec) 在一年内保持稳定. 然而,一个电表仪.
科学领域:
- 医学物理 医学物理
- 仪器科学 仪器科学
背景情况:
- 电表是标准剂量计系统的关键组成部分.
- 准确的校准系数 (kelec) 对于可靠的测量至关重要.
研究的目的:
- 为了评估电表校准系数 (kelec) 的年度变化.
- 为了识别电表性能中的潜在未检测到的错误.
主要方法:
- 研究了两种电表类型:速率模式和整合模式.
- 每月进行电流测量,使用充电发生器的恒定电荷.
- 校准系数被确定在低和中等范围的多范围设备.
主要成果:
- 几乎所有的kelec测量都保持在初始值的0.2%以内.
- 集成模式电表的低范围显示季节性变化超过0.2%.
结论:
- 电表可以产生年度检查无法检测到的错误.
- 用充电发生器进行常规质量保证对于检测微妙的电表漂移至关重要.
相关概念视频
Potentiometer
624
Voltage and current measurements using a standard voltmeter and ammeter alter the circuit being measured either by drawing or resisting the current flow, which introduces uncertainties in the measurements. Null measurements balance the voltages so that no current flows through the measuring device and, therefore, no alterations occur in the measured circuit.
Suppose the emf of a battery needs to be measured. If the battery is directly connected to a standard voltmeter, the measured quantity is...
Suppose the emf of a battery needs to be measured. If the battery is directly connected to a standard voltmeter, the measured quantity is...
624
Biological Effects of Radiation
15.4K
All radioactive nuclides emit high-energy particles or electromagnetic waves. When this radiation encounters living cells, it can cause heating, break chemical bonds, or ionize molecules. The most serious biological damage results when these radioactive emissions fragment or ionize molecules. For example, α and β particles emitted from nuclear decay reactions possess much higher energies than ordinary chemical bond energies. When these particles strike and penetrate matter, they...
15.4K
Galvanometer
2.1K
Common devices, including car instrument panels, battery chargers, and inexpensive electrical instruments, measure potential difference (voltage), current, or resistance using a d'Arsonval galvanometer. This electromechanical instrument is also known as a moving coil galvanometer.
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform...
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform...
2.1K
Coulometry: Overview
1.2K
Coulometry is one of the rapid, most accurate, and precise analytical techniques that determine the quantity of an analyte by measuring the electrical charge needed for its complete electrolysis without using any analytical standards. The total charge passed during electrolysis correlates with the analyte amount by Faraday's laws of electrolysis. For accurate coulometric measurements, a charge equal to Faraday's constant multiplied by the number of electrons involved in the relevant...
1.2K