用数字流体平衡监控技术对现实生活经验的评估
Lisbeth R Leinum1,2, Anders O Baandrup3, Ismail Gögenur2,4
1Department of Urology, Zealand University Hospital, Køge, Denmark.
概括
一种用于流体平衡监测的新数字技术显示出了希望,但需要改进. 虽然对于尿液和静脉注射液体是准确的,但由于发现的错误,口服液体测量组件需要修改.
科学领域:
- 生物医学工程 生物医学工程
- 医疗信息学 医疗信息学
- 临床监测 临床监测
背景情况:
- 医疗保健技术创新可以解决诸如流体平衡监测等挑战.
- 准确的液体平衡图表对于患者护理至关重要.
研究的目的:
- 评估LICENSE数字流体平衡技术的现实世界功能和准确性.
- 将数字流体平衡监测与传统手动文档方法进行比较.
主要方法:
- 许可证系统无线收集了尿液,口腔和静脉液体摄入和输出的数据.
- 来自LICENSE系统的数据与护理人员记录的手动测量数据进行了比较.
- 总共有55名需要流体平衡图表的泌尿病患者被观察到.
主要成果:
- 在数字和手动方法之间的总流体平衡中,观察到的平均差异为-44.2毫升.
- 超过500毫升的显著差异发生在57.4%的病例中.
- 不准确或遗漏的手册文档是主要的错误来源,尽管也注意到了口头许可设备错误.
结论:
- 当正确使用时,LICENSE系统在监测尿液和静脉注射液体方面表现出令人满意的性能.
- LICENSE系统的口服液体测量组件需要进一步开发和修订.
- 数字健康技术为改善流体平衡监测提供了潜力,但需要严格的验证.
相关概念视频
Equipments Used To Measure Blood Pressure
913
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
913
Errors occurring during blood pressure monitoring
667
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
667
Pulse rhythm
777
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
777
Assessing Blood pressure using a doppler ultrasound
1.3K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.3K
Special considerations while measuring blood pressure
711
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
711