手动同步低能量冲击阻抗作为在皮下ICD植入过程中成功进行除测试的预测因素
Naiara Calvo1,2,3, Carlos Rubén López-Perales1,2, Teresa Olóriz1,2
1Arrhytmia Unit, Cardiology Department, Hospital Miguel Servet, Zaragoza, Spain.
Pacing and clinical electrophysiology : PACE
|August 5, 2024
概括
低能同步冲击 (LESS) 阻抗可以预测皮下植入式心脏转换器-除器 (S-ICD) 植入期间的成功除测试 (DFT). 这种方法避免诱导心室动 (VF),并且在预测DFT成功方面具有很高的准确性.
科学领域:
- 心脏病学 心脏病学
- 医疗器械 医疗器械
- 电子生理学 电子生理学
背景情况:
- 在皮下 implantable cardioverter-defibrillator (S-ICD) 植入过程中建议进行除检测 (DFT).
- 冲击阻抗受植入物位置,患者因素和设备特征的影响,会影响除成功.
- 低能同步冲击 (LESS) 是一种评估冲击阻抗而不诱导心室动 (VF) 的方法.
研究的目的:
- 为了比较从LESS和DFT获得的阻抗值.
- 评估 LESS阻抗在预测S-ICD植入过程中成功的DFT的诊断准确度.
主要方法:
- 一项研究包括连续的S-ICD植入.
- 冲击阻抗使用对联t测试进行了比较.
- 后勤回归和ROC曲线分析评估了成功DFT的LESS阻抗的预测准确性.
主要成果:
- LESS 和 DFT 的阻抗值显示出很高的相关性 (r2 = 0.97).
- 失败的DFT患者的BMI更高,LESS和DFT阻抗值明显更高.
- ROC分析表明,在预测成功的DFT时,LESS阻抗 (AUC84%) 的诊断性能很好,截止值<94Ω.
结论:
- 术内LESS阻抗测量可以可靠地预测S-ICD植入期间成功的DFT.
- 这种方法避免了诱导VF的需要,简化了S-ICD植入程序.
相关概念视频
Three-Phase Short Circuit—Unloaded Synchronous Machine
Conducting a three-phase short circuit test on an unloaded synchronous machine helps understand its impact on the system. The AC fault current's oscillogram, with the DC offset removed, reveals that the waveform amplitude decreases from an initially high value to a steady-state level for one phase of the machine.
This behavior occurs due to the magnetic flux produced by the short-circuit armature currents. Initially, these currents follow high-reluctance paths but eventually shift to...
This behavior occurs due to the magnetic flux produced by the short-circuit armature currents. Initially, these currents follow high-reluctance paths but eventually shift to...
Cardiopulmonary Resuscitation III: AED Use
Introduction to AEDAn Automated External Defibrillator (AED) is a portable medical device that analyzes the heart's rhythm and, if necessary, delivers an electrical shock to help the heart re-establish an effective rhythm during sudden cardiac arrest (SCA). SCA occurs when the heart suddenly and unexpectedly stops beating, leading to a loss of blood flow to the brain and other vital organs. In such emergencies, time is of the essence, and using an AED, combined with Cardiopulmonary...