气溶采样技术的方法开发和验证,用于分析电子烟气溶中的尼古丁
Maarten Dill1, Eric Deconinck1, Sophia Barhdadi1
1Sciensano, Scientific Direction Chemical and Physical Health Risks, Service of Medicines and Health Products, J. Wytsmanstraat 14, B-1050 Brussels, Belgium.
Molecules (Basel, Switzerland)
|August 10, 2024
概括
一种新方法准确地测量电子烟气溶中的尼古丁,这对于确保产品安全和质量至关重要. 这种经过验证的方法支持国家卫生当局监测日益增长的电子烟市场.
科学领域:
- 分析化学 分析化学
- 环境健康 环境健康
背景情况:
- 电子烟的普及需要对安全性和质量进行市场监测.
- 欧盟烟草制品指令要求对电子烟产品进行排放研究.
- 由于电子烟排放没有经过验证的方法,这阻碍了可靠的分析.
研究的目的:
- 开发和验证一种用于确定电子烟气溶液中的尼古丁的方法.
- 确保电子烟排放测量的准确性和可重复性.
- 为监管合规提供可靠的分析方法.
主要方法:
- 使用商业蒸汽装置生成电子烟气溶.
- 在剑桥过器上收集了气溶.
- 在提取后使用超高性能液体染色学与二极管阵列检测 (UHPLC-DAD) 量化尼古丁.
- 使用总错误方法和准确性配置文件验证了方法.
主要成果:
- 准确性概况证实了方法的有效性,β-预期耐受性间隔在±20%的接受限度内.
- 可重复性和中间精度是可以接受的,最高相对标准偏差 (RSD) 低于5%.
- 经过验证的方法已成功应用于分析15种商业电子液体中的尼古丁.
结论:
- 实现了电子烟排放分析方法的完整验证,重点是尼古丁的确定.
- 开发的方法确保了对监管监督至关重要的可靠测量.
- 这种系统方法可以适应验证其他电子烟排放分析方法.
更多相关视频
09:30A Microcontroller Operated Device for the Generation of Liquid Extracts from Conventional Cigarette Smoke and Electronic Cigarette Aerosol
Published on: January 18, 2018
8.3K
08:33Use of Capillary Aerosol Generator in Continuous Production of Controlled Aerosol for Non-Clinical Studies
Published on: April 12, 2022
3.4K
相关概念视频
Sampling Methods: Sample Types
196
Sampling materials are classified into three main types: solid, liquid, and gas.
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
196
Atomic Absorption Spectroscopy: Atomization Methods
391
Atomic Absorption Spectroscopy (AAS) atomizes samples through flame atomization or electrothermal atomization. Flame atomization typically involves a nebulizer and spray chamber assembly to combine the sample with a fuel–oxidant mixture, creating a fine aerosol mist that enters a burner. Typically, the fuel and oxidant are combined in an approximately stoichiometric ratio. However, for atoms that are easily oxidized, a fuel-rich mixture may be more advantageous. Only about 5% of the...
391