单打开和关闭和连续脉冲式流动测试仪的比较研究
Hao Wang1, Zhiqian Lu2, Zhongxi Zhou2,3
1School of Electrical and Information Engineering, Jiangsu University, Zhenjiang, Jiangsu Province, 212013, P. R. China.
Cardiovascular engineering and technology
|August 19, 2024
概括
单一的开闭试验器 (SOCVT) 与连续脉冲式流动试验器 (CPFVT) 相比,在关闭体积分数上没有统计差异. 这表明SOCVT可能是一个可行的替代品对关节检测.
科学领域:
- 生物医学工程 生物医学工程
- 心血管设备测试测试器
- 血液动力学 血液动力学
背景情况:
- 准确评估中枢功能对于患者的治疗结果至关重要.
- 现有的门测试方法在模拟生理条件方面存在局限性.
- 为了可靠的设备评估,需要对体外试验进行标准化.
研究的目的:
- 为了比较单个开闭试验器 (SOCVT) 和连续脉动流动试验器 (CPFVT) 之间的闭合体积分量的测量.
- 使用这两种仪器,评估使用这两种仪器进行的中枢检测的准确性和潜在差异.
主要方法:
- 进行了一项比较研究,使用了四种 mitra 状况 (正常,40%扩张,60%扩张,剪切修复).
- 血液动力学条件是根据ISO 5840标准选择的.
- 在SOCVT和CPFVT中测量了闭合体积分数,并计算了16个案例的错误.
主要成果:
- 连续脉动流动测试仪 (CPFVT) 的流速波形更多地受到血液动力学条件的影响,而不是形态.
- 12个案例在闭合体积分数中显示出10-20%的错误,而3个案例显示出2.2-5.5%的错误.
- 在所有测试的中枢条件中,在CPFVT和SOCVT之间的闭合体积分数中没有发现统计学上显著的差异 (P > 0.05).
结论:
- 尽管存在一些测量错误,但SOCVT和CPFVT对中枢的闭合体积分数结果相似.
- 使用典型的中枢流波形标准化的SOCVT测试可以提高其实用性.
- SOCVT 提供了一种潜在的替代性对关节检测,提供了一种简化方法.
相关概念视频
Pipe Flowrate Measurement
622
In pipe flow measurement, orifice, nozzle, and Venturi meters are commonly used to determine fluid flowrates by constricting the flow area, which increases fluid velocity and reduces pressure. This pressure difference, governed by Bernoulli's principle and adjusted for real-world conditions, is essential for calculating flowrate. Each meter type is suited to specific applications based on accuracy, efficiency, and compatibility with various flow conditions.
The orifice meter is a simple,...
The orifice meter is a simple,...
622
Typical Model Studies
352
Fluid mechanics model studies often utilize scaled-down systems to predict fluid behavior in full-scale environments, such as river flows, dam spillways, and structures interacting with open surfaces. Maintaining Froude number similarity in river models is crucial, as it replicates surface flow features like wave patterns and velocities.
352
Rapidly Varying Flow
56
Rapidly varying flow (RVF) in open channels is characterized by abrupt changes in flow depth over a short distance, with the rate of depth change relative to distance often approaching unity. These flows are inherently complex due to their transient and multi-dimensional nature, making exact analysis difficult. However, approximate solutions using simplified models provide valuable insights into their behavior.Key Features of Rapidly Varying FlowRVF is commonly observed in scenarios involving...
56
Flow Table Test
139
The flow table test is an established method used to assess the workability of concrete, particularly useful for evaluating highly flowable concrete mixes. This test employs an apparatus that consists of a wooden board topped with a steel plate, collectively weighing 35 pounds. The board is connected to a base via a hinge and measures 27.6 inches on each side.
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
139
Heart Valves
4.4K
The human heart is a complex organ with an intricate system of valves that regulate blood flow. There are two main types of valves: atrioventricular (AV) valves and semilunar valves.
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
4.4K