达到20万m/z及以上:原生电子捕获 电荷减少 质谱学 解卷 大型生物制药分析物中的异质信号
Kyle I P Le Huray1, Tobias P Wörner2, Tiago Moreira1
1Astbury Centre for Structural Molecular Biology, School of Molecular and Cellular Biology, Faculty of Biological Sciences, University of Leeds, Woodhouse Lane, Leeds LS2 9JT, U.K.
ACS central science
|September 2, 2024
概括
电子捕获电荷减小 (ECCR) 通过减少离子电荷状态来增强原生质谱. 这种技术成功地从异质样本 (如病毒囊) 中解复杂的信号,从而使更大的生物分子得到分析.
科学领域:
- 生物化学 生物化学
- 分析化学 分析化学
- 质谱测量质量谱测量
背景情况:
- 原生质谱在检测大型生物分子方面表现出色,但由于重叠的信号,与异质样本作斗争.
- 在异质样本中复杂的离子分布对准确的表征提出了重大挑战.
研究的目的:
- 应用电子捕获电荷减值 (ECCR) 来改进大型异质生物分子组件的特性.
- 扩大用于Orbitrap质谱分析的质量与电荷 (m/z) 比率上限.
主要方法:
- 原生ECCR是在细菌的Chaperonin GroEL和腺相关病毒 (AAV) 囊组件上进行的.
- 分析使用Q Exactive UHMR质谱仪进行,仪器进行了轻微修改.
主要成果:
- ECCR显著降低了AAV8体的电荷状态,高达90%,转移信号超过10万m/z.
- 对于高度异质的AAV样本,包括带有遗传载荷的样本,实现了电荷状态分辨率和平均质量确定.
- UHMR仪器检测到超过20万m/z的减电离子,证明了前所未有的m/z上层检测极限.
结论:
- 在原生质谱学中,ECCR是解异质信号的强大工具.
- 这项研究证明了 Orbitrap 仪器记录的最高 m/z 信号,扩大了它对更重的分析物的适用性.
相关概念视频
Mass Spectrometry: Overview
4.7K
Mass spectrometry is an analytical technique used to determine the molecular mass and molecular formula of a compound. The basic principle of mass spectrometry is to generate ions from the analyte molecule and measure these ion abundances against their molecular mass. One common type of ionization, known as electrospray ionization or EI, bombards the analyte molecules in the gas phase with high-energy electron beams. The electron beams displace an electron from the molecule and leave...
4.7K
Mass Spectrometers
5.3K
This lesson details the instrumentation of a mass spectrometer—a physical instrument to perform mass spectrometry on analyte molecules and record the characteristic mass spectra. This is achieved via three chief functions:
5.3K
Electrospray Ionization (ESI) Mass Spectrometry
751
Higher molecular weight biomolecules are nonvolatile compounds that may decompose before ionizing or vaporizing during mass analysis with conventional electron impact ionization methods. Accordingly, electrospray ionization (ESI) is the favored method for vaporizing and ionizing biomolecules as it circumvents rapid fragmentation and enables the recording of mass signals for the entire biomolecule.
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
751
Mass Spectrometry: Complex Analysis
736
Mass spectrometry is an important technique for the identification of pure compounds. However, it has some limitations for the analysis of complex mixtures, often due to excessive fragmentation making the spectrum too complicated to decipher. Mass spectrometry can be combined with suitable separation methods in sequence, forming hyphenated methods, which are useful in the analysis of complex mixtures.
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
736
Mass Analyzers: Common Types
579
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
579
High-Resolution Mass Spectrometry (HRMS)
1.3K
The resolution of a mass spectrometer depends on the efficiency of separating ions with different ion masses. The mass of an atom is approximated to the sum of the masses of protons and neutrons inside, considering the masses of protons and neutrons as equal. However, the masses of the proton (1.6726 × 10−24 g) and neutron (1.6749 × 10−24 g) are not truly equal. There is a minor error in the expression of atomic masses relative to the simplest atom of hydrogen. For...
1.3K