:

Goro Nishimura1, Takahiro Suzuki2, Yukio Yamada3

  • 1Hokkaido University, Research Institute for Electronic Science, Sapporo, Japan.

PubMed
概括

在组织中优化光检测需要了解噪声. 这项研究模拟了深度检测极限,为改善临床应用,如吸入风险评估,找到10-15毫米的最佳源探测器距离.

相关概念视频

Super-resolution Fluorescence Microscopy01:37

Super-resolution Fluorescence Microscopy

Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
6.9K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
6.0K
Total Internal Reflection Fluorescence Microscopy01:05

Total Internal Reflection Fluorescence Microscopy

Total internal reflection fluorescence microscopy or TIRF is an advanced microscopic technique used to visualize fluorophores in samples close to a solid surface with a higher refractive index, such as a glass coverslip. TIRF only allows fluorophores in proximity to the solid surface to be excited. When light from a medium with a lower refractive index (such as air) hits the glass coverslip at a critical angle, the light undergoes total internal reflection stead of passing through the glass.
5.7K