血压估计基于动脉反射波的时间由振荡计脉冲波分解
IEEE journal of biomedical and health informatics
|September 5, 2024
概括
一种新的单一传感器方法通过分析波组件,使用脉冲传输时间 (PTT) 准确估计血压 (BP). 这种创新方法简化了BP测量,提供与多传感器技术相比的精确结果.
科学领域:
- 生物医学工程 生物医学工程
- 心血管生理学心血管生理学
- 医疗器械 医疗器械
背景情况:
- 传统的振荡计血压 (BP) 估计可能受到脉冲振幅损害的限制.
- 现有的脉冲过渡时间 (PTT) 方法用于BP估计,通常需要多个传感器,增加复杂性.
- 基于PTT的BP估计的创新为提高准确性提供了潜力,特别是在具有挑战性的生理条件下.
研究的目的:
- 引入和验证一种新的单传感器方法,用于使用脉冲过渡时间 (PTT) 估计振荡计血压 (BP).
- 通过将振荡计脉冲分解为向前和反射波组件来简化PTT测量.
- 根据既定标准和数据集,评估拟议的单传感器PTT方法的准确性和精度.
主要方法:
- 开发了一种使用单个传感器进行振荡计血压估计的新方法.
- 通过分析振荡计脉冲中前进和反射波之间的时间间隔来简化PTT测量.
- 使用两个数据集验证了该方法:150个手臂振荡计记录和425个健康个体的手腕振荡计记录.
主要成果:
- 该研究确定了与缩压和腹缩压相关的特定波间.
- 达到最小标准偏差误差:手臂7.03 mmHg (心缩) 和5.96 mmHg (心径缩),手腕7.98 mmHg (心缩) 和5.39 mmHg (心径缩).
- 在数据集中表现出0.01 mmHg的缩和腹缩血压的平均误差,符合ANSI/AAMI/ISO81060-2标准.
结论:
- 新的单传感器PTT方法提供了准确和精确的振荡计血压估计.
- 这种简化方法与既定方法相比显示出更高的性能,并符合国际标准.
- 建议对各种人群进行进一步验证,包括那些患有肥胖,动脉硬或心房动的人群.
相关概念视频
Sites for measruring blood pressure
1.5K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
The Brachial Artery: Primary Site for Blood Pressure Measurement
1.5K
Assessment of blood pressure in brachial artery(two-step method)
661
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
661
Special considerations while measuring blood pressure
710
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
710
Assessment of blood pressure in brachial artery(one-step method)
560
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
560
Measurement of Blood Pressure
896
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
896
Assessing Blood pressure using a doppler ultrasound
1.2K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.2K