概括
这项研究引入了小型化现场自我校准 (MISSC) 温度计,用于可靠的长期温度测量. 开发的MISSC温度计使用和-固定点,可在难以进入的地方进行准确的校准.
科学领域:
- 计量学 计量学 计量学
- 材料科学 材料科学 材料科学
- 测温仪测温仪是指温度测量仪.
背景情况:
- 确保长期温度测量的准确性是一项挑战,特别是当温度计无法进行常规校准时.
- 为偏远或难以进入的环境开发强大可靠的温度测量解决方案至关重要.
研究的目的:
- 开发和验证一个小型化的现场自校准 (MISSC) 温度计.
- 使用 (Ga) 和- (Ga-Zn) 固定点进行准确的温度尺度转移.
- 为了在传统校准不切实际的环境中实现可靠的长期温度监测.
主要方法:
- 设计了一个MISSC温度计,包括Ga和Ga-Zn固点电池,T型热电偶和薄膜电阻加热.
- 开发了一种测量和校准 (M&C) 系统,其中包括温度测量,数据处理和迷你动力加热控制的子系统.
- 通过从室温控制加热并将相位过渡高原与标准固定点进行比较,实施了现场自校准过程.
主要成果:
- 证明了MISSC温度计拟议的硬件设计的可行性.
- 使用Ga和Ga-Zn固定点验证了现场自我校准方法的有效性.
- 确定了一个 Φ16 mm × 25 mm MISSC 温度计作为能够可靠自我校准的最微型设备.
结论:
- 开发的MISSC温度计及其自我校准方法对于准确的温度测量是有效的.
- 具有现场自校准能力的小型化温度计为在具有挑战性的环境中长期监测提供了可行的解决方案.
- 这项技术提高了温度测量的可靠性,在应用程序中可访问性有限.
相关概念视频
Equipments Used to Measure Body Temperature
980
Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
980
Temperature Measurement Sites
1.6K
A thermometer measures body temperature. The common sites for measuring body temperature are the oral cavity, axillary region, temporal artery, and skin surface, such as the forehead, abdomen, and axilla. True core body temperature is assessed in the rectum, tympanic membrane, pulmonary artery, esophagus, and urinary bladder.
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
1.6K