用于X射线成像水晶光谱仪在位波长校准的X射线源
K Shah1, L Delgado-Aparicio1, B F Kraus1
1Princeton Plasma Physics Laboratory, Princeton, New Jersey 08543, USA.
The Review of scientific instruments
|September 30, 2024
概括
精确的等离子流速测量需要在现场校准X射线成像水晶光谱仪 (XICS). 这项研究开发了X射线源来监测晶体中的温度变化,防止对多普勒变化的误解.
科学领域:
- 等离子体物理学的物理学
- 射线光谱 X射线光谱
- 材料科学 材料科学 材料科学
背景情况:
- 射线成像水晶光谱仪 (XICS) 对于融合装置中的等离子体诊断至关重要.
- 晶体温度的变化会导致光谱线的变化,模仿多普勒变化,导致不准确的等离子体流速测量.
- 在现场校准对于纠正这些由温度引起的变化至关重要.
研究的目的:
- 开发和评估X光源,用于XICS的现场校准.
- 测量X射线光谱线的温度依赖的变化.
- 为了确定晶格格子常数的热膨胀,以准确的波长校准.
主要方法:
- 实验室使用铜 (Cu) 射线源,加热的石英 (102) 晶体和皮拉图斯探测器进行实验.
- 测量Cu Kα1和Kα2的直线偏移是结晶温度的函数.
- 对布拉格反射平面与温度的二维格子常数的评估.
主要成果:
- 对Cu Kα1和Kα2X射线的量化温度依赖转移.
- 确定了晶体温度与2D网格常数之间的关系.
- 验证了使用X射线源在现场校准XICS的可行性.
结论:
- 发达的X射线源对于在核聚变装置中XICS的准确现场校准至关重要.
- 考虑到温度依赖的变化对于避免等离子体流速测量的错误至关重要.
- 开发的方法提供了在动态实验环境中精确波长校准的途径.
相关概念视频
X-ray Crystallography
23.8K
The size of the unit cell and the arrangement of atoms in a crystal may be determined from measurements of the diffraction of X-rays by the crystal, termed X-ray crystallography.
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
23.8K
Atomic Absorption Spectroscopy: Radiation and Light Sources
359
Atomic absorption spectroscopy (AAS) relies on the Beer-Lambert law, which requires that the radiation source emits a narrow range of wavelengths to match the absorption characteristics of the analyte atom. The primary criteria for choosing an appropriate radiation source in AAS is to provide a precise and intense emission at specific wavelengths that will allow accurate detection of the analyte.
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
359
Atomic Absorption Spectroscopy: Instrumentation
578
An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.
The atomizer used in AAS can be either a flame atomizer or an...
The atomizer used in AAS can be either a flame atomizer or an...
578
IR Spectrometers
1.1K
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
1.1K
UV–Vis Spectrometers
1.3K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
1.3K
X-ray Diffraction of Biological Samples
3.8K
X-ray diffraction or XRD is an analytical tool that utilizes X-rays to study ordered structures such as crystalline organic and inorganic samples, polycrystalline materials, proteins, carbohydrates, and drugs.
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
3.8K