自动图像转录用于使用移动卫生技术进行围产期血压监测
Nasim Katebi1,2, Whitney Bremer1, Tony Nguyen1
1Department of Biomedical Informatics, Emory University, Atlanta, Georgia, United States of America.
PLOS digital health
|October 2, 2024
概括
本研究介绍了用于血压 (BP) 设备的自动图像转录方法,改进了远程监控的数据传输,特别是在低识字人口中. 深度学习模型准确地转录了BP读数,有助于怀孕和产后护理.
科学领域:
- 生物医学工程 生物医学工程
- 医疗信息学 医疗信息学
- 机器学习 机器学习
背景情况:
- 自测血压 (BP) 监测在将数据传输到临床记录方面面临挑战,特别是在低识字人口中.
- 将振荡计设备的数据集成到电子健康记录中,需要强大的方法来准确地读取转录.
- 移动健康 (mHealth) 计划需要有效的方式来收集和利用来自不同人群的BP数据,包括那些资源较少的环境.
研究的目的:
- 开发和验证一种自动化图像转录技术,用于从振荡计装置读取血压.
- 提高自测血压数据的可访问性和可用性,用于临床监测,特别是在怀孕和产后.
- 促进BP数据与医疗健康记录的整合,以改善在低识字环境中的患者管理.
主要方法:
- 开发了一个深度学习模型,将LCD本地化的You Only Look Once (YOLO) 和数字识别的卷积神经网络结合起来.
- 包括色彩校正和值在内的图像处理技术被应用以减轻反射和文物.
- 该模型在危地马拉和格鲁吉亚的mHealth研究数据集上进行了训练和测试,涉及孕妇和产后妇女.
主要成果:
- 自动化图像转录模型的平均绝对误差 (MAE) 远低于FDA推的5mmHg.
- 与没有转移学习的模型相比,具有转移学习的设备独立模型和设备特定模型表现出优异的性能.
- 缩和腹缩血压的特定MAE值为1.2和0.8mmHg (格鲁吉亚IMPROVE) 和0.9和0.5mmHg (危地马拉第二组).
结论:
- 拟议的自动化图像转录模型是有效的,适合一般使用准确的BP设备.
- 这项技术可以显著改善远程BP监测和数据集成,特别有利于低识字和低资源人口.
- 该方法支持通过可访问的mHealth数据加强怀孕期间和产后高血压的管理.
相关概念视频
Equipments Used To Measure Blood Pressure
838
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
838
Errors occurring during blood pressure monitoring
607
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Several factors...
607
Assessment of blood pressure in brachial artery(one-step method)
556
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
556
Measurement of Blood Pressure
891
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
891
Assessing Blood pressure using a doppler ultrasound
1.2K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.2K
Special considerations while measuring blood pressure
708
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
708