传感器位置的微小变化会导致心脏输出读数的系统变化:对临床实践的影响
Caroline Dinesen1,2, Simon Tilma Vistisen2, Peter Juhl-Olsen3,4
1Department of Cardiothoracic- and Vascular Surgery, Aarhus University Hospital, Anesthesia Section, Palle Juul-Jensens Boulevard 99, Aarhus N, 8200, Denmark.
Journal of clinical monitoring and computing
|October 10, 2024
概括
在传感器位置的小变化显著改变心脏输出 (CO) 测量使用第4代FloTrac软件. 精确的传感器放置对于在重症监护机构的精确血液动力学监测至关重要.
科学领域:
- 麻醉学 麻醉学
- 关键护理医学 关键护理医学
- 心血管生理学心血管生理学
背景情况:
- 精确的血液动力学监测对于心脏手术后的患者管理至关重要.
- 该FloTrac系统 (第四代) 提供连续的心力输出 (CO) 估计.
- 传感器的定位是影响血液动力学测量的已知因素.
研究的目的:
- 系统地评估微小的垂直传感器位置变化对CO和其他血液动力学变量的影响,这些变量由第四代FloTrac软件测量.
- 为了确定FloTrac CO测量对传感器高度变化的灵敏度.
- 在血液动力学监测中评估这些位置变化的临床意义.
主要方法:
- 一项前性研究,涉及20名心脏手术后的患者.
- 用第四代FloTrac测量心脏输出 (CO),平均动脉压 (MAP),全身血管阻力 (SVR) 和中风体积变化 (SVV).
- 传感器被随机放置在中间轴平面上,比中间轴平面高4厘米或低4厘米.
主要成果:
- 传感器位置的4厘米垂直移动导致了CO的显著变化 (P < 0.001).
- 转换器位置较高时,CO增加了大约4.1%,转换器位置较低时减少了3.4%,平均每厘米约为~1%.
- 随着传感器位置的变化,MAP和SVR也发生了显著的变化 (P < 0.001),而SVV没有受到影响 (P = 0.98).
结论:
- 垂直传感器位置显著影响使用第四代FloTrac软件获得的心脏输出测量.
- 即使是传感器高度的小 (4厘米) 变化也可能导致血液动力学数据的临床显著变化.
- 根据FloTrac数据,对传感器水平和患者定位的细致关注对于可靠的目标导向治疗至关重要.
相关概念视频
Special considerations while measuring blood pressure
707
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
707
Equipments Used To Measure Blood Pressure
837
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
837
Assessing Blood pressure using a doppler ultrasound
1.2K
To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.2K
Pre-Procedural Guidelines for Assessing Blood Pressure
522
Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
522
Imaging Studies for Cardiovascular System I:Echocardiography
299
Cardiac imaging studies encompass a wide range of noninvasive and minimally invasive techniques designed to visualize the heart's structure and function in detail. One such technique is echocardiography, which uses high-frequency ultrasound waves to produce detailed images of the heart, known as echocardiograms.
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
299
Pulse
783
The pulse is one of the most fundamental physiological indicators of the body's cardiovascular health. It is the rhythmic expansion and contraction of the arterial walls in response to the pressure generated by the heart's pumping action.
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
783