在血液透析溶液中开发,验证和应用Na,K,Ca和Mg的快速序列方法,由HR-CS F AAS在血液透析溶液中进行
Flávia E Schwartzhaupt1, Leandro Kolling2,3, Wiliam Boschetti4
1Federal University of Rio Grande do Sul, Post-graduation Program in Pharmaceutical Sciences, Faculty of Pharmacy, Av. Ipiranga, 2752, Porto Alegre, RS 90610-000, Brazil.
Journal of AOAC International
|October 15, 2024
概括
一种新方法使用高分辨率连续源火焰原子吸收光谱学 (HR-CS F AAS) 准确确定血液透析溶液中的,,和. 这种方法大大减少了质量控制的分析时间.
科学领域:
- 分析化学 分析化学
- 频谱学是一种光谱学.
背景情况:
- 血液透析溶液对于脏替代疗法至关重要.
- 准确量化这些溶液中的电解质对于患者的安全至关重要.
研究的目的:
- 开发和验证一种用于测定血液透析溶液中的 (Na), (K), (Ca) 和 (Mg) 的新方法.
- 使用高分辨率连续源火焰原子吸收光谱学 (HR-CS F AAS) 结合实验设计 (DOE) 进行方法优化.
主要方法:
- 使用Doehlert矩阵和可取性函数优化关键变量 (乙烯流量,燃烧器高度,Cs和La度).
- 根据国际协调理事会 (ICH) 准则进行验证,评估选择性,线性,精度,准确性和稳定性.
主要成果:
- 精确 (RSD <2.7%) 和准确 (恢复100.5103.0%) 的Na,K,Ca和Mg的确定.
- 已建立的校准曲线在每个元素的特定度范围内.
- 在单个分析运行中成功进行了连续的多元件确定.
结论:
- 开发的HR-CS F AAS方法为血液透析溶液中的电解质分析提供了一种经过验证,高效和准确的方法.
- 与现有技术相比,这种方法可显著减少样品和标准准备时间.
- 介绍了一个新的替代品,目前的欧洲药典方法用于血液透析溶液分析.
相关概念视频
Atomic Absorption Spectroscopy: Lab
311
For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
311
Dialysis
603
Dialysis is a diffusion-based purification process that separates analyte molecules from a complex matrix. This is accomplished by allowing molecules in the solution to pass through a semipermeable membrane into a liquid on the other side. The membrane is usually made of cellulose acetate or cellulose nitrate, and the second liquid must be miscible with the solution. Ions (e.g., chloride or sodium) or organic molecules (e.g., glucose) can pass through the membrane pores, which generally have...
603
Hemodialysis II: Procedure and Complications
2
DialyzersA hemodialysis (HD) dialyzer is a plastic cartridge containing thousands of parallel hollow fibers, which serve as semipermeable membranes. These fibers are typically made from cellulose-based or other synthetic materials. During HD, blood is pumped into the top of the cartridge and distributed among these fibers. Simultaneously, dialysis fluid, known as dialysate, is introduced into the bottom of the cartridge, bathing the outside of the fibers. Across the semipermeable membrane,...
2
Hemodialysis III: Nursing Management
2
The nursing management of a patient undergoing hemodialysis includes several critical steps, starting with a thorough assessment before the procedure.Before the Hemodialysis ProcedureFirst, record the patient's vital signs—blood pressure, heart rate, respiratory rate, and temperature—to establish a baseline. This baseline is essential for detecting conditions such as hypotension that could impact the patient's response to dialysis. Document the patient's pre-dialysis weight, as this...
2
Qualitative Analysis
22.0K
For solutions containing mixtures of different cations, the identity of each cation can be determined by qualitative analysis. This technique involves a series of selective precipitations with different chemical reagents, each reaction producing a characteristic precipitate for a specific group of cations. Metal ions within a group are further separated by varying the pH, heating the mixture to redissolve a precipitate, or adding other reagents to form complex ions.
For instance, group IV...
For instance, group IV...
22.0K
Capillary Electrophoresis: Applications
343
Capillary electrophoretic separations offer various modes, each with unique applications. These modes include capillary zone electrophoresis, capillary gel electrophoresis, capillary array electrophoresis, capillary isoelectric focusing, capillary isotachophoresis, micellar electrokinetic chromatography, and capillary electrochromatography.
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
343