气体纯度对使用感应合等离子体双重质谱学观察到的NO反应性的影响
Amanda D French1, Kirby P Hobbs1, Richard M Cox1
1Pacific Northwest National Laboratory, Richland, WA, 99352, USA. amanda.french@pnnl.gov.
The Analyst
|October 17, 2024
概括
使用纯化氧化 (NO) 在感应合等离子体双重质谱法 (ICP-MS/MS) 中显著减少干扰,并增强信号强度,以改进元素分析.
科学领域:
- 分析化学 分析化学
- 质谱测量质量谱测量
背景情况:
- 在ICP-MS/MS中,干扰去除主要依赖于碰撞/反应电池气体选择.
- 反应气体杂质对ICP-MS/MS光谱的影响仍未得到充分研究.
研究的目的:
- 评估净化氧化 (NO) 对ICP-MS/MS中干扰去除和光谱质量的影响.
- 评估NO气体中的水 (H2O) 杂质对元素分析的影响.
主要方法:
- 评估了60个元素的反应性,使用99.5%的NO,带有和没有净化器,以达到<100 pptV H2O.
- 实验重点是V,Ce,Tl和Th,使用不同流速的净化NO (0.22-1.49毫升min-1).
主要成果:
- 净化NO大大减轻了氧化物干扰.
- 总的离子灵敏度得到了改善,特别是在更高的气体流速下.
- 产品分布转向有利,对主要预期产品的信号增强了多达一个数量级.
结论:
- 减少NO气中的H2O杂质可以提高ICP-MS/MS的性能.
- 增强的信号和减少的背景可以在各种ICP-MS/MS分析应用中取得进展.
相关概念视频
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
407
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
407
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
657
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
657
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
535
Inductively coupled plasma (ICP) is the most widely used plasma source in atomic emission spectroscopy (AES), also known as Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES). The ICP source, or torch, consists of three concentric quartz tubes with argon gas flowing through them. A spark from a Tesla coil initiates the ionization of argon, generating a high-temperature plasma.
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
535
Gas Chromatography: Types of Detectors-II
338
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
338
Atomic Emission Spectroscopy: Lab
149
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
149
Atomic Emission Spectroscopy: Interference
175
In atomic emission spectroscopy (AES), high-temperature atomizers excite a broad range of elements and molecules that generate complex emissions from sources such as oxides, hydroxides, and flame combustion products in the flame or plasma. Several strategies can be employed to minimize spectral interferences caused by overlapping emission lines or bands. These include increasing instrument resolution, choosing alternative emission lines, optimally placing the detector in low-background regions,...
175