光极化对激光光斑对比成像的影响,用于微血管流动的定制幻影
Nasrin Amini1, Ali Esteki1, Mohsen Ahmadi1
1Department of Medical Physics & Biomedical Engineering, School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
Scientific reports
|November 4, 2024
概括
极化控制显著改善激光光斑对比成像 (LSCI) 用于血液流量测量. 使用垂直偏振增强信号噪声比,并减少微循环成像中的错误.
科学领域:
- 生物医学光学 生物医学光学
- 医疗成像医学成像
- 流体动力学 流体动力学
背景情况:
- 激光光斑对比成像 (LSCI) 是血液流量评估的重要非侵入性技术.
- LSCI性能对光散射和系统参数敏感,影响信号噪声比 (SNR).
- 光的极化为提高LSCI能力提供了一个有希望的途径.
研究的目的:
- 研究光极化操纵对LSCI流量测量性能的影响.
- 评估极化控制在毛细血管内微循环成像中的有效性.
- 确定最佳的极化状态,以提高LSCI中的精度和SNR.
主要方法:
- 开发了一个定制的幻影,用各种直径 (1.1-2.8毫米) 的管嵌入在模拟组织矩阵中.
- 操纵事件和反射光的偏振 (平行,垂直,没有偏振).
- 比较LSCI系统在不同极化条件下检测不同尺寸和深度管道的流量时的性能.
主要成果:
- 深度显著影响LSCI流量检测灵敏度.
- 发生光和反射光之间的垂直偏振产生了最小的测量误差.
- 垂直偏振也导致了较高的SNR与平行偏振和没有偏振控制相比.
结论:
- 极化操纵是优化LSCI性能的一个关键因素.
- 垂直偏振显著提高LSCI准确性和SNR,特别是对于微循环.
- 这种技术有可能在临床应用中实现更可靠的血液流动成像.
相关概念视频
Phase Contrast and Differential Interference Contrast Microscopy
7.5K
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
7.5K
Confocal Fluorescence Microscopy
13.1K
Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
13.1K