儿童自动运动血压测量:一项初步研究
Frank A Treiber1, Francis McCaffrey2, William B Strong1
1Dept. of Pediatrics and the Georgia Institute for the Prevention of Human Disease and Accidents, Medical College of Georgia, Augusta, GA 30912.
Pediatric exercise science
|November 13, 2024
概括
昆顿410自动监视器准确地测量运动期间儿童的心率. 它为群体比较提供了可接受的血压估计值,尽管腹压读数略低于传统方法.
科学领域:
- 儿童心脏病学 儿童心脏病学
- 运动生理学 运动生理学
- 生物医学工程 生物医学工程
背景情况:
- 在运动期间精确监测儿童的心血管参数对于健康评估至关重要.
- 自动化设备为高效和客观的测量提供了潜力.
- 对新技术与既定方法进行验证是必不可少的.
研究的目的:
- 为了评估Quinton 410自动运动监测器对儿童血压和心率的准确性.
- 为了比较昆顿410的读数与同时的听力和心电图测量.
- 评估各种条件的设备性能,包括休息,姿势变化和运动.
主要方法:
- 同时测量血压和心率,使用昆顿410,听觉和心电图.
- 在休息,姿势变化,跑步机运动工作负载和积极恢复期间收集数据.
- 在一组幼儿 (平均年龄为7.9岁) 中分析了测量一致性和相关性.
主要成果:
- 在所有条件下,Quinton 410在心率测量方面表现出高精度 (平均差异<1.0 bpm).
- 缩血压读数显示出良好的相关性和与听觉相似性,除了在初始运动期间.
- 透气血压读数略低于听觉读数 (平均平均差异 = 3.1 mmHg).
结论:
- 昆顿410在各种活动期间为儿童提供高度准确的心率读数.
- 该设备在运动期间为儿科患者群体的群体比较提供了可接受的血压估计.
- 对于精确的个人腹压评估,可能需要进一步的验证.
相关概念视频
Assessment of blood pressure in brachial artery(two-step method)
656
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
656
Pre-Procedural Guidelines for Assessing Blood Pressure
522
Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the...
522
Assessment of blood pressure in brachial artery(one-step method)
556
This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
Prepare for the Procedure:
556
Measurement of Blood Pressure
880
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
880
Equipments Used To Measure Blood Pressure
835
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
835
Special considerations while measuring blood pressure
705
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
705