圆束计算机断层扫描机的灰色值和噪声行为 - - 一项体外研究
Nicolly Oliveira-Santos1,2, Hugo Gaêta-Araujo3, Rubens Spin-Neto4
1Department of Oral Surgery and Stomatology, Division of Diagnostic Sciences, School of Dental Medicine, University of Bern, Bern, 3010, Switzerland.
Dento maxillo facial radiology
|November 20, 2024
概括
圆束计算机断层扫描 (CBCT) 图像显示不同机器和采集设置的平均灰色值 (MGV) 和噪声的显著变化. 视野大小是影响MGV的主要因素,特别是骨密度.
科学领域:
- 放射学和成像科学 放射学和成像科学
- 医学物理 医学物理
- 牙科成像技术 牙科成像技术
背景情况:
- 圆束计算机断层扫描 (CBCT) 对于牙科和口面部成像至关重要.
- 标准化质量保证对于可靠的CBCT成像至关重要.
- 了解获取参数对图像质量指标的影响至关重要.
研究的目的:
- 系统地评估CBCT成像中的平均灰度值 (MGVs) 和噪声.
- 评估各种采集参数对图像质量的影响.
- 为了比较9台不同的CBCT机器的性能.
主要方法:
- 使用DIN6868-161幻影与骨,软组织和空气等价物.
- 在9台CBCT机器上扫描了幻影,改变了管电流 (mA),管电压 (kV),视野 (FOV) 和旋转角度.
- 使用Kruskal Wallis和Dunn-Bonferroni测试分析了MGV和对比噪声指标 (CNI).
主要成果:
- 管道电流一般不会影响MGVs.
- 在一些机器 (Viso G7,Veraview X800) 中,增加kV降低了MGVs.
- 在大多数机器中,MGV随着FOV高度的增加而减少;旋转角度的影响最小.
- 随着辐射的减少和更大的FOVs,CNI较低.
结论:
- 在MGV和噪音方面存在显著的机器间变化.
- FOV大小是影响MGV的主要因素,特别是在骨等价放射密度方面.
- 图像质量指标通常随着获取参数的增加而下降.
相关概念视频
Computed Tomography
4.3K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
4.3K
X-ray Imaging
5.4K
German physicist Wilhelm Röntgen (1845–1923) was experimenting with electrical current when he discovered that a mysterious and invisible "ray" would pass through his flesh but leave an outline of his bones on a screen coated with a metal compound. In 1895, Röntgen made the first durable record of the internal parts of a living human: an "X-ray" image (as it came to be called) of his wife’s hand. Scientists worldwide quickly began their own experiments with...
5.4K