离子风对冠状放电离子移动性光谱学样本分布的影响
Jiyao Wang1, Chong Zhang1, Wenyan Shi1
1Division of Advanced Manufacturing, Tsinghua Shenzhen International Graduate School, Shenzhen, 518055, China.
Talanta
|November 27, 2024
概括
这项研究优化了冠状放电离子流动性光谱法 (IMS) 中的离子风,以更好地分析样本. 轴向注射方法提高了检测化学品的灵敏度,并确保了电池的安全性.
科学领域:
- 分析化学 分析化学
- 化学物理 化学物理
- 工具分析 工具分析
背景情况:
- 冠状放电 (CD) 的离子风会影响离子移动谱仪 (IMS) 中的空气流.
- 这种离子风对样本分布的影响经常被忽视.
- 平衡反应物离子密度和离子风强度对于防止样品稀释至关重要.
研究的目的:
- 调查离子风对CD-IMS.样本分布的影响.
- 开发一种方法来减轻由离子风引起的样本稀释.
- 为了提高CD-IMS的灵敏度和恢复性能.
主要方法:
- 提出了一个轴向注射模式,使用空心针进行样品引入.
- 评估了离子风对空气流和样本分布的影响.
- 量化了二甲基甲基酸盐 (DMMP) 和电池电解质组件的检测极限.
主要成果:
- 轴向注射模式显著提高了灵敏度和恢复性能.
- 对DMMP单体的检测极限从15ppbv降至5ppbv.
- 成功检测到电池电解质泄漏的低度 (10 mg/L或1 nL碳酸盐在5秒内).
结论:
- 拟议的轴向注射技术有效地解决了CD-IMS中的离子风诱导的样品稀释.
- 这种方法为快速和敏感的监控应用提供了途径,特别是电池安全.
- 优化离子风管理是推进IMS技术的关键.
相关概念视频
Electrospray Ionization (ESI) Mass Spectrometry
735
Higher molecular weight biomolecules are nonvolatile compounds that may decompose before ionizing or vaporizing during mass analysis with conventional electron impact ionization methods. Accordingly, electrospray ionization (ESI) is the favored method for vaporizing and ionizing biomolecules as it circumvents rapid fragmentation and enables the recording of mass signals for the entire biomolecule.
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
735
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
646
In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
646
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
525
Inductively coupled plasma (ICP) is the most widely used plasma source in atomic emission spectroscopy (AES), also known as Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES). The ICP source, or torch, consists of three concentric quartz tubes with argon gas flowing through them. A spark from a Tesla coil initiates the ionization of argon, generating a high-temperature plasma.
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
525
Chemical Ionization (CI) Mass Spectrometry
690
The molecular ion peak of a molecule in the mass spectrum provides vital information for molecular identification. However, conventional electron impact ionization can lead to the rapid dissociation of some molecular ions before they reach the detector. A milder ionization method is required to increase the lifetime of such ionized analyte molecules. Chemical ionization (CI) is a gas-phase protonation reaction useful for mass-analyzing analyte molecules that are easily protonated to yield the...
690
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
395
Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
395
Mass Spectrometers
5.2K
This lesson details the instrumentation of a mass spectrometer—a physical instrument to perform mass spectrometry on analyte molecules and record the characteristic mass spectra. This is achieved via three chief functions:
5.2K