一个护士发明的尿液排水袋固定装置的可行性测试
Karen Meade1, Bridgette Kreuz, Loraine Sinnott
1Author Affiliations: Clinical Nurse Specialist (Ms Meade), Gynecology Oncology & Urology, James Cancer Hospital and Solove Research Institute, The Ohio State University Wexner Medical Center; Occupational Therapist (Dr Kreuz), OhioHealth; Consulting Research Statistician (Dr Sinnott), Data Coordination and Analysis Center, College of Optometry, The Ohio State University; Clinical Quality and Patient Safety Coordinator (Ms Kuhn), James Cancer Hospital and Solove Research Institute; Clinical Nurse Specialist-Acute Care Nursing (Dr Potts) and Clinical Nurse Specialist-Surgical (Ms Ring), The Ohio State University Wexner Medical Center East Hospital; and Nurse Consultant (Dr Brown), Zenith Healthcare Solutions, Columbus, Ohio.
膀下方袋 (B3) Buddy设备非常可接受,可用于在患者出行期间固定尿液排水袋. 这种由护士发明的设备有望改善患者的移动性和护理.
科学领域:
- 护理创新 护理创新
- 医疗设备的可用性 医疗设备的可用性
- 患者移动辅助器 患者移动辅助器
背景情况:
- 排尿袋可以阻碍患者的行走.
- 需要安全装置来促进安全的患者移动.
- 护士领导的发明可以解决临床挑战.
研究的目的:
- 为了评估囊下膀 (B3) 的可行性,Buddy.
- 评估B3 Buddy用于固定尿液排水袋的可用性和可接受性.
- 为了确定设备在病人行走期间的有效性.
主要方法:
- 一项涉及患者和临床医生的干预研究.
- 参与者在B3 Buddy的住院环境中至少行走了90英尺.
- 经过验证的工具 (实施措施的可接受性,系统可用性尺度) 用于数据收集.
主要成果:
- 所有21名患者都认为B3Buddy可接受 (95%CI:83.9%-100%),超过了70%的门.
- 临床医生报告了高可用性,系统可用性尺度 (SUS) 的平均得分为90.1 (95%CI:84.9-95.3).
- 有学士学位的临床医生和注册护士的SUS分数较高.
结论:
- 对于最终用户来说,B3 Buddy设备是可以接受和使用的.
- 这些发现支持B3 Buddy在临床应用中的可行性.
- 发明者正在寻求制造合作伙伴关系以进一步发展.