相关概念视频

Computed Tomography01:10

Computed Tomography

4.3K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
4.3K

您也可能阅读

相关文章

通过共同作者、期刊和引用图与本文相关的文章。

排序
Same author

The radiographic bubbly fecal pattern of intestinal pneumatosis in newborns revisited.

Pediatric radiology·2026
Same author

Radiomics-informed CT protocol optimization through feature-level trade-off analysis in a phantom study.

Physics in medicine and biology·2026
Same author

Imaging Pitfalls in Surface Bone Lesions: Florid Reactive Periostitis Mimicking Parosteal Osteosarcoma - A Multimodality Case Series.

RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin·2026
Same author

The impact of increasing femoral width on anatomic risk factors for patellar instability in paediatric patients: a retrospective MRI case-control study.

BMC musculoskeletal disorders·2026
Same author

[Does endoprosthetic treatment of medial femoral neck fractures in obese patients using a minimally invasive anterolateral approach have a higher complication rate? : A retrospective case-control study].

Orthopadie (Heidelberg, Germany)·2026
Same author

Knowledge preservation in the era of big science and AI: strategies for sustainable scientific research.

Nature communications·2026
Same journal

Cumulative radiation exposure from repeated CT imaging in hospitalised COVID-19 patients: Does it matter?

European journal of radiology open·2026
Same journal

Prostate cancer: Value of the PI-FAB score and volume-adjusted PSA density in recurrence assessment after HIFU.

European journal of radiology open·2026
Same journal

Performance of magnetic-resonance imaging radiomics in prediction of response after neoadjuvant chemotherapy in head and neck squamous cell carcinoma: A systematic review and meta-analysis.

European journal of radiology open·2026
Same journal

Ultra-low contrast agent dosage in photon-counting CT angiography of the thoracoabdominal aorta.

European journal of radiology open·2026
Same journal

Intratumoral and peritumoral radiomics for preoperative prediction of pathological complete response to neoadjuvant immunochemotherapy in patients with esophageal squamous cell carcinoma.

European journal of radiology open·2026
Same journal

Role of apparent diffusion coefficient (ADC) and MRI-derived parameters in identifying p53-abnormal subtypes of endometrial cancer.

European journal of radiology open·2026
查看所有相关文章

相关实验视频

Updated: Jun 5, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

14.4K

低剂量肺CT:优化诊断辐射剂量 - 一个幻影研究

Michael Zellner1,2, Sebastian Tschauner1,2,3, Mathias S Weyland4,5

  • 1University Children's Hospital Zürich, Department of Diagnostic Imaging, Zurich, Switzerland.

European journal of radiology open
|December 10, 2024
PubMed
概括
此摘要是机器生成的。

这项研究开发了一种定量方法,用图像清晰度来评估低剂量肺部CT图像质量. 在显著降低辐射剂量时,可以达到诊断质量,这表明在不牺牲准确性的情况下,可以降低剂量.

关键词:
诊断成像,肺部成像,肺部成像低剂量的低剂量辐射辐射辐射辐射辐射辐射辐射断层扫描,X射线计算机扫描

更多相关视频

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
08:17

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy

Published on: June 7, 2015

15.7K
Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
06:33

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement

Published on: July 29, 2013

11.3K

相关实验视频

Last Updated: Jun 5, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

14.4K
Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
08:17

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy

Published on: June 7, 2015

15.7K
Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
06:33

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement

Published on: July 29, 2013

11.3K

科学领域:

  • 放射学 放射学是一门学科.
  • 医疗成像医学成像
  • 辐射剂量计 辐射剂量计

背景情况:

  • 在低剂量肺部计算机断层扫描 (CT) 中评估图像质量对于诊断准确性至关重要.
  • 需要定量方法来确定产生诊断图像的最低辐射暴露.

研究的目的:

  • 建立一个量化方法来评估低剂量肺CT图像质量.
  • 为了确定肺部CT诊断图像质量所需的最小辐射剂量.

主要方法:

  • 轴体积肺CT扫描在成人和儿童幻体上进行,剂量不同.
  • 图像的清晰度是通过测量在肺-多叶膜接口的西格形曲线的度来量化.
  • 放射科医生使用利克特尺度评估诊断图像质量,以确定适当图像的清晰度切线.

主要成果:

  • 诊断图像质量在713 HU/mm (成年幻影) 的中位斜率下实现,有效剂量为0.13 mSv.
  • 对于儿科幻影,诊断质量达到1139HU/mm,有效剂量为0.08mSv.
  • 这些剂量对应于CTDIvol值分别为0.14 mGy和0.13 mGy.

结论:

  • 图像清晰度测量提供了对肺部CT图像质量的可靠的定量评估.
  • 低剂量的肺CT可以在有效辐射剂量为0.13mSv (成人) 和0.08mSv (儿科) 的情况下达到足够的诊断质量.
  • 在肺CT中大幅减少辐射剂量是可行的,而不会影响诊断准确度.