一种用于放射治疗CT图像的新型可变形图像注册软件的基准测试和性能评估
Shorug S Alshammari1,2, Sridhar Yaddanapudi1, Blaž Kušnik3
1Department of Radiation Oncology, Mayo Clinic, Jacksonville, FL, USA.
Technical innovations & patient support in radiation oncology
|December 24, 2024
概括
一个新的可变形图像注册 (DIR) 软件DirOne显示出临床上可接受的准确性. 它的性能与商业系统相提并论,支持其在放射治疗规划和图像细分方面的应用.
科学领域:
- 医疗成像医学成像
- 辐射疗法 辐射疗法
- 图像注册 图像注册 图像注册
背景情况:
- 可变形图像注册 (DIR) 对于精确的放射治疗至关重要.
- 对新的DIR软件与已建立的商业系统进行评估对于临床采用至关重要.
研究的目的:
- 为了评估和比较一个新的DIR软件,DirOne.
- 为了比较DirOne的性能与商业系统MIM和VelocityAI.
- 评估是否符合AAPM任务组132 (TG-132) 的指导方针.
主要方法:
- 使用了三个公共数据集:男性骨盆幻影,4DCT胸部扫描 (DIR-Lab) 和头部和部CT (DIREP).
- 使用目标注册错误 (TRE),子相似系数 (DSC) 和平均距离达成协议 (MDA) 作为评估指标.
- 通过使用已建立的数据集和指标,将DirOne与MIM和VelocityAI进行基准比较.
主要成果:
- DirOne的平均TRE值为3.3 ± 3.1毫米,与MIM (2.7 ± 3.7毫米) 和VelocityAI (3.4 ± 2.4毫米) 相比.
- 在前两个数据集中,DirOne的DSC为0.96 ± 0.02,略低于MIM和VelocityAI (0.98 ± 0.02和0.98 ± 0.01,分别).
- 在DIREP数据集中,DirOne的MDA值为0.73 ± 0.34和DSC值为0.91 ± 0.03.
结论:
- 新的DIR软件DirOne显示出临床上可以接受的准确性.
- DirOne的性能与商业DIR系统相美.
- DirOne显示了辐射疗法应用的潜力,包括细分和剂量扭曲.
更多相关视频
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
972
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
496
相关概念视频
Computed Tomography
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...